83
TONE?
Tone Enable Query
Syntax:
TONE?
Parameters:
None
Function:
This query returns a value indicating whether the tone feature is not available,
enabled or disabled.
Returns:
<state>
<state> is a <number> which can be -1, 0 or 1. If -1 then the tone feature is not
available for the detector in use. If 0 then the tone feature is disabled. If 1
then the tone feature is enabled.
Related Commands:
TONE
TRIGOUT
Trigger Out Mode Select
Syntax:
TRIGOUT <mode>
Parameters:
The parameter <mode> is of type <string>. The valid values for this parameter
are listed below.
“CMPLT”
specifies conversion complete mode
“FREQ”
specifies frequency mode
“LVL”
specifies level comparator mode
“INTG”
specifies integration level mode
“OFF”
specifies off mode
Function:
This command determines in which mode the trigger out output on the back
panel will operate.
Certain modes require additional parameters to be set.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTINTG?, TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL,
TRIGOUTPOL?
TRIGOUT?
Trigger Out Mode Query
Syntax:
TRIGOUT?
Parameters:
None
Function:
This query returns the current mode of the trigger out output.
Returns:
<mode>
<mode> is of type <string> in the double quote format.
Related Commands:
TRIGOUT, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTINTG?, TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL,
TRIGOUTPOL?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com