72
LLO?
Local Lockout Enable Query
Syntax:
LLO?
Parameters:
None
Function:
The query returns the state of the local lockout.
Returns:
<state>
<state> is of type <number> and represents the integer 0, if local lockout is
disabled, or 1, if local lockout is enabled.
Related Commands:
LLO
MODE_n
Acquisition Mode Select
Syntax:
MODE_n <mode>
where n is either A or B
Parameters:
The parameter <mode> is of type <string>. Its range depends on the detector
attached to the meter. The valid values for this parameter are listed below.
“DCSNGL”
specifies DC single mode
“DCCONT”
specifies DC continuous mode
“INTG”
specifies integration mode
“PPSNGL”
specifies P-P single mode
“PPCONT”
specifies P-P continuous mode
“SNGLPULSE”
specifies single pulse mode
“CONTPULSE”
specifies continuous pulse mode
Function:
This command sets the mode to be used to acquiring subsequent readings.
Note that the current units will be changed automatically if they are not
compatible with the new mode.
Related Commands:
MODE_n?, UNITS_n, UNITS_n?
MODE_n?
Acquisition Mode Query
Syntax:
MODE_n?
where n is either A or B
Parameters:
None
Function:
This query returns the acquisition mode currently being used by the meter.
Returns:
<mode>
<mode> is of type <string> in the double quote format.
Related Commands:
MODE_n, UNITS_n, UNITS_n?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com