69
EXT?
External Trigger Enable Query
Syntax:
EXT?
Parameters:
None
Function:
This query returns a value showing whether the external trigger input is
enabled or disabled.
Returns:
<state>
<state> is of type <number> that represents the integer 0, if the external
trigger input is disabled, or 1, if the external trigger is enabled.
Related Commands:
EXT, EXTEDGE, EXTEDGE?
EXTEDGE
External Trigger Edge Select
Syntax:
EXTEDGE <edge>
Parameters:
The parameter <edge> is either 0 or 1. If <edge> is 0 then the external trigger
is defined as a falling edge. If <edge> is 1 then the external trigger is defined as
a rising edge.
Function:
This command defines whether the external trigger input on the back panel is
falling edge or rising edge active.
Related Commands:
EXT, EXT?, EXTEDGE
EXTEDGE?
External Trigger Edge Query
Syntax:
EXTEDGE?
Parameters:
None
Function:
This query returns a value showing which external trigger edge is the active
edge.
Returns:
<edge>
<edge> is of type <number> and represents 0, if the external trigger is defined
as a falling edge, or 1, if the external trigger is defined as a rising edge.
Related Commands:
EXT, EXT?, EXTEDGE