89
USRREF_n?
User Defined Reference Value Query
Syntax:
USRREF_n?
where n is either A or B
Parameters:
None
Function:
This query returns the user defined reference value. This value will be in
watts for Low-Power and High-Power detectors and joules for Energy detec-
tors.
Returns:
<refval>
<refval> is of type <number> in exponential notation.
Related Commands:
REFSEL_n, REFSEL_n?, STOREF_n, USRREF_n
USRRESP_n
User Responsivity Factor Select
Syntax:
USRRESP_n <val>
where n is either A or B
Parameters:
The parameter <val> is of type <number>. The units of this parameter match
the responsivity units of the detector in use.
Function:
This command is used to define a responsivity factor that will be used by the
2835-C in measurement calculationwhen user calibration is enabled.
Related Commands:
ATTN_n, ATTN_n?, LAMBDA_n, LAMBDA_n?, RESP_n?,
USRCAL_n, USRCAL_n?, USRRESP_n?
USRRESP_n?
User Responsivity Factor Query
Syntax:
USRRESP_n?
where n is either A or B
Parameters:
None
Function:
This query returns the user defined responsivity factor that will be used by
the 2835-C in measurement calculation when user calibration is enabled.
Returns:
<val>
<val> is of type <number> and represents a floating point number. The units
of <val> match the responsivity units of detector in use.
Related Commands:
ATTN_n, ATTN_n?, LAMBDA_n, LAMBDA_n?, RESP_n?,
USRCAL_n, USRCAL_n?, USRRESP_n
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com