87
UNITS_n
Units Select
Syntax:
UNITS_n <units>
where n is either A or B
Parameters:
<units> is of type <string>. All possible values of units are listed below:
“V”
specifies volts
“A”
specifies amps
“W”
specifies watts
“W/cm”
specifies watts/cm
2
“dBm”
specifies dBm
“dB”
specifies dB
“REL”
specifies linear ratio
“J”
specifies joules
“J/cm”
specifies joules/cm
2
“ERG”
specifies ergs
“E/cm”
specifies ergs/cm
2
A subset of these units will be valid for a particular detector and acquisition mode.
Function:
Sets the units to be used for subsequent measurements. Once the units have
been set, all new readings are given in the new units.
Related Commands:
MODE_n, MODE_n?, UNITS_n?
UNITS_n?
Units Query
Syntax:
UNITS_n?
where n is either A or B
Parameters:
None
Function:
This query returns the units of readings currently being taken.
Returns:
<units>
<units> is of type <string> in the double quote format. The possible values of
units are: “A”, “V”, “W”, “W/cm”, “dBm”, “dB”, “REL”, “J”, “J/cm”, “ERG”, or “E/
CM” depending on the detector and acquisition mode.
Related Commands:
MODE_n, MODE_n?, UNITS_n
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com