61
DETSW_n
Detector Switch Setting
Syntax:
DETSW_n <setting>
where n is either A or B
Parameter:
<setting> is of type <string> in a double quoted format. The possible values of
<setting> are “S” for short, “I” for intermediate, and “L” for long.
Function:
The 818J-S10 and the 818J-09 have a switch setting for short, intermediate and
long energy pulse measurement. This command must be used to inform the
2835-C what the switch is set to when using these detectors.
If the attached detector doesn’t support a switch then this command will
cause an execution error.
Related Commands:
DETSW_n?
DETSW_n?
Detector Switch Setting Query
Syntax:
DETSW_n?
where n is either A or B
Parameter:
None
Function:
This query returns the switch setting assumed by the 2835-C. This setting
must match the actual switch setting on the detector or readings will be
inaccurate.
Returns:
<setting>
<setting> is of type <string> in the double quote format. The possible values
of <setting> are “S” for short, “I” for intermediate, and “L” for long. If the
attached detector doesn’t support a switch the return value will be “N/A”.
Related Commands:
DETSW_n
DISP
Display Brightness
Syntax:
DISP <level>
Parameters:
The parameter <level> is of type <number> which rounds to an integer in the
range 0 to 3. 0 corresponds to off, 1 to dim, 2 to normal, and 3 to bright.
Function:
This command sets the brightness of the display to a desired level. The valid
levels are off, dim, normal, and bright.
Returns:
None
Related Commands:
DISP?