60
CH?
Available Channel Query
Syntax:
CH?
Parameters:
None
Function:
This query is used to see which channels are available in the meter. A chan-
nel is available if a calibration module was plugged into that channel on
powerup and passed the self test.
Returns:
<channels>
<channels> is of type <string> in the double quote format. The possible values
of <string> are:
“A” - only channel A is available
“B” - only channel B is available
“AB” - Both channel A and B are available
Related Commands:
None
DETMODEL_n?
Detector Model Query
Syntax:
DETMODEL_n?
where n is either A or B
Parameters:
None
Function:
This query returns the model number of the detector that belongs to the
memory module attached to the specified channel.
Returns:
<model>
<model> is of type <string> in the double quote format.
Related Commands:
ATTNSN_n?, CALDATE_n?, DETSN_n?
DETSN_n?
Detector Serial Number Query
Syntax:
DETSN_n?
where n is either A or B
Parameter:
None
Function:
This query returns the serial number of the detector that belongs to the
memory module attached to the specified channel.
This serial number must match the serial number of the detector attached to
the 2835-C.
Returns:
<detSN>
<detSN> is of type <string> in the double quote format.
Related Commands:
ATTNSN_n?, CALDATE_n?, DETMODEL_n?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com