24
WARNING
This product is equipped with a 3 wire grounding type plug. Any inter-
ruption of the grounding connection can create an electric shock hazard.
If you are unable to insert the plug into your wall plug receptacle,
contact your electrician to perform the necessary alterations to assure
that the green (green-yellow) wire is attached to earth ground.
2.6
Detector Connection and Setup
Connect the detector to its calibration module as shown, Figure 6. The
detector’s model and serial numbers should match the model and serial
numbers found on the calibration module. Insert the calibration module into
the front panel input port of the Model 2835-C. An alignment pin forces the
proper orientation of the calibration module.
Detector Cable BNC
Calibration Module
Figure 6 – Connecting a Detector with its Calibration Module
Plug calibration
modules into these ports.
INPUT
CH B
INPUT
CH A
CH A
CH B
R/S A
R/S B
Figure 7 – Model 2835-C Detector Calibration Module Input Port
NOTE
If the detector being connected is a battery powered Newport Energy
Detector then be sure to configure the DET SWITCH POS menu function to
match the position of the switch on the detector, see Section 2.4.12. The
2835-C defaults to the intermediate switchposition, I.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com