ii. Configure the trigger output to TRIG AT INTG and set the energy level at
which you wish the exposure to stop, see Section 2.4.7. Adjust the polarity
of the trigger output to open the shutter when the trigger output is active.
iii. Set the meter to INTG mode with AUTO on and then press the R/S key to
begin the integration and open the exposure controlling shutter.
iv. The display value will increase until the INTG trigger output threshold is
crossed. At the threshold, integration will stop and the trigger output will
close the shutter. The display now shows the final integration value from
the exposure.
NOTE
You can also configure the Model 2835-C as an exposure timer-controller by
configuring the trigger output to TRIG AT FREQ and setting the trigger
frequency appropriately.