85
TRIGOUTINTG?
Trigger Out Integration Level Query
Syntax:
TRIGOUTINTG?
Parameters:
None
Function:
This query returns the trigger out integration energy level at which the trigger
out will become inactive if using the trigger out in integration acquisition
mode.
Returns:
<level>
<level> is of type <number> in exponent notation with units of joules.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL, TRIGOUTPOL?
TRIGOUTLVL
Trigger Out Level Select
Syntax:
TRIGOUTLVL <level>
Parameters:
The parameter <level> is of type <number>. The units of this parameter are
Watts for Low-Power and High-Power detectors and Joules for Energy detec-
tors.
Function:
This command is used to set the level used by the trigger out level compara-
tor mode. When the power or energy level is at or above this level the trigger
out will go to its active state.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTINTG?, TRIGOUTLVL?, TRIGOUTPOL, TRIGOUTPOL?
TRIGOUTLVL?
Trigger Out Level Query
Syntax:
TRIGOUTLVL?
Parameters:
None
Function:
This query returns the trigger out level used by the trigger out level compara-
tor mode.
Returns:
<level>
<level> is of type <number> in exponent notation with units of Watts for Low-
Power and High-Power detectors and Joules for Energy detectors.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTINTG, TRIGOUTINTG?, TRIGOUTLVL, TRIGOUTPOL, TRIGOUTPOL?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com