Performing a Test
56
EasyOne Pro™ / EasyOne Pro™ LAB
3.4 Vital Capacity (SVC)
Slow vital capacity measurements determine the vital capacity, starting
from tidal breathing. (You can configure the device for direct measurement
of the vital capacity, i.e., without initial tidal breathing (see
page 120). For this test type, too, the ATPS values (
A
mbient
T
emperature
P
ressure
S
aturated = spirometer conditions) must be converted to BTPS val-
ues (
B
ody
T
emperature
P
ressure
S
aturated = body conditions) (see
After three acceptable trials, the message
Session complete!
appears.
Select the patient (see
Prepare the patient (see
on page 46) and explain the test
maneuvers to him or her:
take the
spirette™
into the mouth with the lips sealing around the
spirette™
, taking care not to block its opening with the tongue or bite
down excessively on the
spirette™
breathe calmly in and out
when the start test prompt appears, take a maximum inspiration
followed by an immediate, but unhurried, exhalation.
Select
SVC
.
The slow vital capacity acquisition screen appears.