EasyOne Pro™ / EasyOne Pro™ LAB
117
System Settings
Note
The predicted normal is corrected by the factor
i
, if no
specific calculation is defined for the ethnic group in the
selected predicted normal publication.
Note
The settings selected here affect the predicted normals, the
interpretation and the displayed parameter readings.
Note
When the "Best Value" option is selected, the values for the
"Best" column are selected as follows:
For the parameters listed below, the best value is the highest
value from all accepted trials:
The IC, IRV, VT and ERV parameters are calculated from the
average of all accepted trials.
For PEF, PIF and the volume parameters, the best value is the
highest value from all accepted trials:
Ratios such as FEV1/FVC, FEF50 / VCmax , MTC1 are
recalculated.
Other parameters are taken from the highest ranked trial
(= 1).
For DLCO tests, the "Best Value" / "Best Trial" selection is
irrelevant. The values are calculated from the average of all
accepted trials.