Performing a Test
48
EasyOne Pro™ / EasyOne Pro™ LAB
controls, such as ventilation, air filtration or ultraviolet
decontamination of air, should be used to prevent disease
transmission.
Risk of Infection—Clean the nose clip after each patient.
Measurement
Insert the new
spirette™
in the sensor.
Attach the nose clip to the patient.
Press
Start
.
The device advises you to keep the
spirette™
closed until the test start
prompt is displayed. This is necessary to set the baseline correctly.
Note
ndd recommends leaving the spirette™ packaging in place
until after the baseline has been set.
Block off the
spirette™
on one end and confirm the message with
OK
.
When the start test prompt is displayed, hand the sensor over to your
patient and instruct him or her to perform the maneuver as explained
earlier.
Note
If the "Manual Test Stop" option was selected in the
configuration (see
must terminate the measurement by clicking the Test End
button or by pressing the space bar or the Enter key.
During the test, you will see the flow-volume loop and the volume-time
curve on the display. All curves and measured values will be displayed after
the test.