EasyOne Pro™ / EasyOne Pro™ LAB
155
Appendix
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
EasyOne Pro™
is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that
EasyOne Pro™
is used in
such an environment.
Immunity Test
IEC 60601 Test Level Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic dis-
charge (ESD) to
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV
± 8 kV
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
to EN 61000-4-4
± 2 kV for power sup-
ply lines
± 1 kV for input/out-
put lines
the product
has no power
cords or input/
output lines
Surge to EN
61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
the product
has no power
cords
Voltage dips,
short interrup-
tions and volt-
age variations on
power supply
input lines to EN
61000-4-11
< 5% U
T
(>95% dip in U
T
) for
0.5 cycles
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
) for
5 s
the product
has no power
cords
Power frequency
(50/60 Hz) mag-
netic field to EN
61000-4-8
3 A/m
passed
Power frequency magnetic
fields should be at levels char-
acteristics of a typical location
in a typical commercial or hos-
pital environment.
NOTE U
T
is the AC mains voltage prior to application of the test level.