EasyOne Pro™ / EasyOne Pro™ LAB
53
Performing a Test
Conduct the tests as described in section
Note
The volume-time curve (Fig. 3-16) does not display the
forced inspiration at the end of the maneuver.
Bronchodilation (
Post
-Tests)
The
Post
test is usually performed to determine the patient's response to
bronchodilating asthma medication. This is done by administering a bron-
chodilator to the patient after he or she has performed a spirometry test
(
Pre
-Test). Approximately 10 to 20 minutes after the medication (when
bronchodilator shows effect, refer to medication labelling) a second spirom-
etry test is performed. After the test the
Pre
and
Post
results are compared.
Post
tests can only be added to previous tests performed on the same day.
Select the patient.
Select the test, then click
Add Post
to close the window.
The window with the patient's existing tests appears. With
left) you display the detailed results of the
Pre
tests (Fig. 3-15 right).
Conduct the tests as described in section
As soon as three acceptable trials have been performed, the message
Ses-
sion complete!
appears and a quality grade from
A
(optimal) to
F
is dis-
"Quality Messages and Quality Grades"
Fig. 3-15
Detailed test results
Summary of Contents for easyone pro
Page 1: ...Version V04b EasyOne Pro EasyOne Pro LAB Operator s Manual ...
Page 133: ...EasyOne Pro EasyOne Pro LAB 133 Specifications Power con sumption max 80 VA Life time 7 years ...