TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8100A with Options 001, 002, or 8100B Digital Voltmeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 2 of 2
4.3
Resistance Measurements
TI Range & Calibrator
(k
)
(k
)
(k
)
4.3.2
1 k
range
1.0000
0.9980
to
1.0020
”
10 k
range
10.000
9.980
to
10.020
”
100 k
range
100.00
99.80
to
100.20
”
1000 k
range
1000.0
998.0
to
1002.0
TI Range & Calibrator
(M
)
(M
)
(M
)
4.3.2
10 M
10.000
9.980
to
10.020
Affix a special calibration label stating:
”All resistance ranges have been calibrated to +0.2% of range.
The V ac frequency response has not been calibrated.”
NAVAIR 17–20VQ–05
42