TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8600A Digital Multimeter (Including Options 001 and/or 002)
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+20 V
10 V x 0
0.000
–0.002
to
0.002
”
+20 V
10 V x .1
1.000
0.998
to
1.002
”
+20 V
10 V x .2
2.000
1.997
to
2.003
”
+20 V
10 V x .3
3.000
2.997
to
3.003
”
+20 V
10 V x .4
4.000
3.996
to
4.004
”
+20 V
10 V x .5
5.000
4.996
to
5.004
”
+20 V
10 V x .6
6.000
5.996
to
6.004
”
+20 V
10 V x .7
7.000
6.995
to
7.005
”
+20 V
10 V x .8
8.000
7.995
to
8.005
”
+20 V
10 V x .9
9.000
8.994
to
9.006
4.1.6
–20 V
10 V x 1
10.000
9.994
to
10.006
4.1.9
2 V
1 V x 1
1.0000
0.9994
to
1.0006
”
200 V
100 V
100.00
99.94
to
100.06
”
1200 V
100 V
100.0
99.8
to
100.2
TI Range Calibrator
mV)
(mV)
(mV)
4.1.9
200 mV
1 V x .1
100.00
99.91
to
100.09
TI 200 mV Range
V)
(
V)
(
V)
4.1.13
10
V LSD
520
510 to 530
4.1.14
10
V LSD
480
470
to
490
4.2
Alternating–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.2.2
2 V
400 Hz–1 V
1.0000
0.9966
to
1.0034
”
20 V
400 Hz–10 V
10.000
9.966
to
10.034
”
20 V
4 kHz–10 V
10.000
9.966
to
10.034
”
20 V
50 kHz–10 V
10.000
9.929
to
10.071
”
200 V
50 kHz–10 V
10.00
9.87
to
10.13
”
200 V
4 kHz–10 V
10.00
9.93
to
10.07
”
200 V
400 Hz–100 V
100.00
99.66
to
100.34
”
1200 V
400 Hz–100 V
100.0
99.2
to
100.8
”
1200 V
4 kHz–10 V
10.0
9.5
to
10.5
NAVAIR 17–20VQ–05
129