TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Cimron 6753 or 6753–1424 Digital Multimeter with the 5775 or 5789 Plug in Unit
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 2 of 2
4.3
Resistance Measurements (5789)
TI Range
Calibrator
(k
)
(k
)
(k
)
4.3.2
1 k
1 k
1.00000
0.99800 to
1.00200
”
10 k
10 k
10.0000
9.9800
to
10.0200
”
100 k
100
k
100.000
99.800
to
100.200
”
1000 k
1 M
1000.00
998.00
to
1002.00
”
10000 k
10 M
10000.0
9980.0
to
10020.0
Affix a special calibration label stating the following information as applicable:
1. The TI direct voltage ratio function has not been calibrated.
2. If the TI consists of a 5789 plug in, the 10 mV and the 100 mV V dc ranges
have not been calibrated; also, all resistace ranges have been calibrated to
+0.2% of range.
3. If the TI consists of a 5775 converter or plug–in, the V ac frequency response,
and measurements above 50 kHz have not been calibrated; also, the 1, 10,
and 100 V V ac ranges have been calibrated to +0.15% of range at 400 Hz.
NAVAIR 17–20VQ–05