TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Cimron DMM 50 Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.0000
–0.0001
to
0.0001
”
+10 V
10 V x .1
1.0000
0.9998
to
1.0002
”
+10 V
10 V x .2
2.0000
1.9997
to
2.0003
”
+10 V
10 V x .3
3.0000
2.9996
to
3.0004
”
+10 V
10 V x .4
4.0000
3.9995
to
4.0005
”
+10 V
10 V x .5
5.0000
4.9994
to
5.0006
”
+10 V
10 V x .6
6.0000
5.9994
to
6.0006
”
+10 V
10 V x .7
7.0000
6.9993
to
7.0007
”
+10 V
10 V x .8
8.0000
7.9992
to
8.0008
”
+10 V
10 V x .9
9.0000
8.9991
to
9.0009
4.1.6
–10 V
10 V x 1
10.0000
0.9990
to
10.0010
TI Range Calibrator
(
V)
(V)
(V)
4.1.9
1 V
1 V x 1
1.00000
0.99990
to
1.00010
”
100 V
100 V
100.000
99.990
to
100.010
”
1 kV
100 V
100.00
99.98
to
100.02
TI 100 mV range
(m
V)
(
m
V)
(
m
V)
4.1.13
1
m
V LSD
502
501
to
503
4.1.14
1
m
V LSD
498
497
to
499
4.2
Alternating–Voltage Measurements
Range
Calibrator
(V)
(V)
(V)
4.2.2
1 V
400 Hz–1 V
1.00000
0.99850
to
1.00150
”
10 V
400 Hz–10 V
10.0000
9.9850
to
10.0150
”
10 V
4 kHz–10 V
10.0000
9.9850
to
10.0150
”
10 V
50 kHz–10 V
10.0000
9.9440
to
10.0560
”
100 V
50 kHz–10 V
10.000
9.872
to
10.128
”
100 V
4 kHz–10 V
10.000
9.967
to
10.033
”
100 V
400 Hz–100 V
100.000
99.850
to
100.150
”
1000 V
400 Hz–100 V
100.00
99.67
to
100.33
NAVAIR 17–20VQ–05
69