TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Monsanto 200A Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 1
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.000
–0.0010
to
0.0010
”
+10 V
10 V x .1
1.000
0.9989
to
1.0011
”
+10 V
10 V x .2
2.000
1.998
to
2.001
”
+10 V
10 V x .3
3.000
2.999
to
3.001
”
+10 V
10 V x .4
4.000
3.999
to
4.001
”
+10 V
10 V x .5
5.000
4.998
to
5.002
”
+10 V
10 V x .6
6.000
5.998
to
6.002
”
+10 V
10 V x .7
7.000
6.998
to
7.002
”
+10 V
10 V x .8
8.000
7.998
to
8.002
”
+10 V
10 V x .9
9.000
8.998
to
9.002
”
+10 V
10 V x 1
10.000
9.998
to
10.002
4.1.6
–10 V
10 V x 1
10.000
9.998
to
10.002
TI Range Calibrator
(
V)
(V)
(V)
4.1.9
100 V
100 V
100.0
99.98
to
100.02
”
1000 V
100 V
100.0
99.9
to
100.1
TI 1 V Range
(m
V)
(
m
V)
(
m
V)
4.1.13
100
m
V LSD 700
600
to
800
4.1.14
100
m
V LSD
200
to
400
NAVAIR 17–20VQ–05
(78 blank) 77