TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8375A Digital Multimeter with Options 003 and/or 007
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 2 of 2
4.2.2
TI Range
Calibrator
(V)
(V)
(V)
”
100 V
50 kHz–10 V
10.000
9.405
to
10.595
”
1000 V
400 Hz–100 V
100.00
99.47
to
100.53
”
1000 V
4 kHz–10 V
10.00
9.59
to
10.41
4.3
Resistance Measurements
TI Range & Calibrator
(
)
(
)
(
)
4.3.2
1000
1000.00
998.00 to
1002.00
”
10000
10000.0
99800
to
10020.0
(k
)
(k
)
(k
)
4.3.2
100 k
100.000
99.800 to
100.200
”
1000 k
1000.00
998.00 to
1002.00
”
10000 k
10000.0
9980.0 to
10020.0
Affix a special calibration label stating:
1. The 0.1 V V dc range has not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
NAVAIR 17–20VQ–05
60