TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Cimron Digital Multimeter, Sanders P/N 3005896G1 as used with 0Q–10/ALM–64*
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL (6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 1
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+
10 V
10 V x 0
0.0000
–0.0002
to
0.0002
”
+
10 V
10 V x .1
1.0000
0.9997
to
1.0003
”
+
10 V
10 V x .2
2.0000
1.9995
to
2.0005
”
+
10 V
10 V x .3
3.0000
2.9992
to
3.0008
”
+
10 V
10 V x .4
4.0000
3.9990
to
4.0010
”
+
10 V
10 V x .5
5.0000
4.9987
to
5.0013
”
+
10 V
10 V x .6
6.0000
5.9985
to
6.0015
”
+
10 V
10 V x .7
7.0000
6.9982
to
7.0018
”
+
10 V
10 V x .8
8.0000
7.9980
to
8.0020
”
+
10 V
10 V x .9
9.0000
8.9977
to
9.0023
”
+
10 V
10 V x 1
10.0000
9.9975
to
10.0025
”
–10 V
10 V x 1
10.0000
9.9975
to
10.0025
”
TI Range
Calibrator
(V)
(V)
(V)
4.1.9
100 V
100 V
100.000
99.975
to 100.025
”
1000 V
100 V
100.00
99.97
to 100.03
TI 10–volt range
(m
V)
(
m
V)
(
m
V)
4.1.13
100
m
V LSD
700
600
to
800
4.1.14
100
m
V LSD
300
200
to
400
*Affix a special calibration label to the TI stating:
“This TI is calibrated in accordance with the usage requirement of 0Q–10/ALM–64”.
NAVAIR 17–20VQ–05
(14 blank) 13