TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8350A Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.0000
–0.0005
to
0.0005
”
+10 V
10 V x .1
1.0000
0.9993
to
1.0007
”
+10 V
10 V x .2
2.0000
1.9992
to
2.0008
”
+10 V
10 V x .3
3.0000
2.9990
to
3.0010
”
+10 V
10 V x .4
4.0000
3.9989
to
4.0011
”
+10 V
10 V x .5
5.0000
4.9987
to
5.0013
”
+10 V
10 V x .6
6.0000
5.9986
to
6.0014
”
+10 V
10 V x .7
7.0000
6.9984
to
7.0016
”
+10 V
10 V x .8
8.0000
7.9983
to
8.0017
”
+10 V
10 V x .9
9.0000
8.9981
to
9.0019
”
+10 V
10 V x 1
10.0000
9.9980
to
10.0020
4.1.6
–10 V
10 V x 1
10.0000
9.9980
to
10.0020
TI Range Calibrator
mV)
(mV)
(mV)
4.1.9
1000 mV
10 V x 1
1000
999.83
to
1000.17
TI Range
Calibrator
(V)
(V)
(V)
4.1.9
100 V
100 V
100.000
99.980
to
100.020
”
1000 V
100 V
100.00
99.93
to
100.07
TI 100 mV Range
V)
(
V)
(
V)
4.1.13
1
V LSD
502
501
to
503
4.1.14
1
V LSD
498
497
to
499
4.2
Alternating–Voltage Measurements
Range
Calibrator
(V)
(V)
(V)
4.2.2
1 V
400 Hz–1 V
1.00000
0.99860
to
1.00140
”
10 V
400 Hz–10 V
10.0000
9.9860
to
10.0140
”
100 V
400 Hz–100 V
100.000
99.860
to
100.140
”
1000 V
400 Hz–100 V
100.00
99.50
to
100.50
”
10 V
4 kHz–10 V
10.0000
9.9860
to
10.0140
”
10 V
50 kHz–10 V
10.0000
9.9460
to
10.0540
NAVAIR 17–20VQ–05
57