TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Hewlett–Packard 3465A Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 2 of 2
4.3
Resistance Measurements
TI Range* & Calibrator
(k
)
(k
)
(k
)
4.3.2
1 k
1.0000
0.9980
to
1.0020
”
10 k
10.000
9.980
to
10.020
”
100 k
100.00
99.80
to
100.20
”
1000 k
1000.0
998.0
to
1002.0
TI Range* & Calibrator
(M
)
(M
)
(M
)
4.3.2
10 M
range
10.000
9.980
to
10.020
Affix a special calibration label stating:
1. The 10 and 100 mV ranges, and the 100 mV V ac range have not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
3. The DC and AC current functions have not been calibrated.
*In more recent models, range sequence are: 2, 20, 200 etc.
NAVAIR 17–20VQ–05
102