TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Cimron 6453A, 6453A–1, 6453A–2, or 6453A–3 Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL (6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 2 of 2
4.3
Resistance Measurements
TI & Calibrator
(k
)
(k
)
(k
)
4.3.2
1 k
range
1.0000
0.9980 to
1.0020
”
10 k
range
10.000
9.980
to
10.020
”
100 k
range
100.00
99.80
to
100.20
”
1000 k
range
1000.0
998.0
to
1002.0
”
10000 k
range
10000
9980
to 10020
Affix a special calibration label stating:
1. The 100 mV V dc range has not been calibrated.
2. All resistance ranges have been calibrated to
0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
NAVAIR 17–20VQ–05