TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Electro Instruments, Inc. 4010 DC Voltmeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL (6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 1
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.000
–0.001
to
0.001
”
+10 V
10 V x .1
1.000
0.999
to
1.001
”
+10 V
10 V x .2
2.000
1.999
to
2.001
”
+10 V
10 V x .3
3.000
2.999
to
3.001
”
+10 V
10 V x .4
4.000
3.999
to
4.001
”
+10 V
10 V x .5
5.000
4.999
to
5.001
”
+10 V
10 V x .6
6.000
5.999
to
6.001
”
+10 V
10 V x .7
7.000
6.999
to
7.001
”
+10 V
10 V x .8
8.000
7.999
to
8.001
”
+10 V
10 V x .9
9.000
8.999
to
9.001
4.1.6
–10 V
10 V x .9
9.000
8.999
to
9.001
TI Range
Calibrator
(V)
(V)
(V)
4.1.9
100 V
100 V
100.00
99.99
to
100.01
”
1000 V
100 V
100.0
99.9
to
100.1
NAVAIR 17–20VQ–05
(22 blank) 21