TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8500A/AEOPT1/2/5/8 as used with the TS 3846/ASM608 Inertial Measuring Unit Test Set
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.00000
–0.0000–1 to
0.0000–1
”
+10 V
10 V x .1
1.00000
0.9991–0
to
1.0009–0
”
+10 V
10 V x .2
2.00000
1.9991–0
to
2.0009–0
”
+10 V
10 V x .3
3.00000
2.9991–0
to
3.0009–0
”
+10 V
10 V x .4
4.00000
3.9991–0
to
4.0009–0
”
+10 V
10 V x .5
5.00000
4.9991–0
to
5.0009–0
”
+10 V
10 V x .6
6.00000
5.9991–0
to
6.0009–0
”
+10 V
10 V x .7
7.00000
6.9991–0
to
7.0009–0
”
+10 V
10 V x .8
8.00000
7.9991–0
to
8.0009–0
”
+10 V
10 V x .9
9.00000
8.9991–0
to
9.0009–0
”
+10 V
10 V x 1
10.00000
9.9991–0
to
10.0009–0
4.1.6
–10 V
10 V x 1
–10.00000
–9.9991–0 to
10.0009–0
4.1.9
1 V
1 V x 1
1.00000
0.9999–1
to
1.0000–9
”
100 V
100 V
100.0000
99.991–0
to
100.009–0
”
1000 V
100 V
100.0000
99.910–0
to
100.090–0
TI 0.1 V Range
V)
(
V)
(
V)
1
V LSD
502
501
to
503
1
V LSD
498
497
to
499
4.2
Alternating–Voltage Measurements (01)
TI Range
Calibrator
(V)
(V)
(V)
4.2.2
1 V
400 Hz–1 V
1.00000
0.99850
to
1.00150
”
10 V
400 Hz–10 V
10.0000
9.9880
to
10.0120
”
100 V
400 Hz–100 V
100.000
99.880
to
100.120
”
1000 V
400 Hz–100 V
100.000
98.200
to
101.800
”
10 V
4 kHz–10 V
10.0000
9.9880
to
10.0120
”
10 V
50 kHz–10 V
10.0000
9.9005
to
10.1005
NAVAIR 17–20VQ–05
151