TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8120A Digital Voltmeter including Options 001 and/or 002
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.000
–0.001
to
0.001
”
+10 V
10 V x .1
1.000
0.999
to
1.001
”
+10 V
10 V x .2
2.000
1.999
to
2.001
”
+10 V
10 V x .3
3.000
2.998
to
3.002
”
+10 V
10 V x .4
4.000
3.998
to
4.002
”
+10 V
10 V x .5
5.000
4.998
to
5.002
”
+10 V
10 V x .6
6.000
5.998
to
6.002
”
+10 V
10 V x .7
7.000
6.998
to
7.002
”
+10 V
10 V x .8
8.000
7.997
to
8.003
”
+10 V
10 V x .9
9.000
8.997
to
9.003
4.1.6
–10 V
10 V x 1
10.000
9.997
to
10.003
TI Range
Calibrator
(V)
(V)
(V)
4.1.9
1 V
1 V x 1
1.0000
0.9997
to
1.0005
”
100 V
100 V
100.00
99.97
to
100.03
”
1000 V
100 V
100.0
99.9
to
100.1
TI 100 mV Range
V)
(
V)
(
V)
4.1.13
10
V LSD
520
510
to
530
4.1.14
10
V LSD
480
470
to
490
4.2
Alternating–Voltage Measurements
Range
Calibrator
(V)
(V)
(V)
4.2.2
1 V
400 Hz–1 V
1.0000
0.9975
to
1.0025
”
10 V
400 Hz–10 V
10.000
9.95
to
10.025
”
10 V
4 kHz–10 V
10.000
9.975
to
10.025
”
100 V
4 kHz–10 V
10.000
9.93
to
10.07
”
100 V
400 Hz–100 V
100.00
99.75
to
100.25
”
1000 V
400 Hz–100 V
100.0
99.3
to
100.7
NAVAIR 17–20VQ–05
93