TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8800A/AA Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 2 of 2
4.3
Resistance Measurements
TI Range
Calibrator
(k
W
)
(k
W
)
(k
W
)
4.3.2
2 k
W
1 k
W
1.00000
0.99800 to
1.00200
”
20 k
W
10
k
W
10.0000
9.9800
to
10.0200
”
200 k
W
100 k
W
100.000
99.800
to
100.200
”
2000 k
W
1 M
W
1000.00
998.00
to
1002.00
TI Range
Calibrator
(M
W
)
(M
W
)
(M
W
)
4.3.2
20 M
W
10 M
W
10.0000
9.9700
to
10.0300
W
W
W
”
200
W
100
W
100.0
99.6
to
100.4
Affix a special calibration label stating:
1. The TI 200 mV V dc range, and the 2 V V ac range frequency response
have not been calibrated.
2. The TI 200
W
, 2 k
W
, 20 k
W, 200
k
W,
and 2000 k
W
resistance ranges have
been calibrated to +0.1% of range.
3. The 20, 200 and 1200 V ac Ranges have been calibrated up to 50 kHz only.
NAVAIR 17–20VQ–05
100