TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Fluke 8010A Digital Multimeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V dc)
(V dc)
(V dc)
4.1.3
+20 V
10 x 0
0.00
–0.01
to
0.01
”
+20 V
10 x .1
1.00
0.98
to
1.02
”
+20 V
10 x .3
3.00
2.98
to
3.02
”
+20 V
10 x .5
5.00
4.98
to
5.02
”
+20 V
10 x .7
7.00
6.98
to
7.02
”
+20 V
10 x .9
9.00
8.98
to
9.02
4.1.6
–20 V
10 x .9
–9.00
–8.98
to
–9.02
TI Range
Calibrator
V dc)
(V dc)
(V dc)
4.1.9
200 mV
1 V x 0.1
100.0
99.8
to
100.2 mV
”
2 V
1 V x 1
1.000
0.998
to
1.002
”
200 V
100 V
100.0
99.8
to
100.2
”
1000 V
100 V
100.0
99.8
to
100.2
TI 200 mV Range
V)
(
V)
(
V)
4.1.13
100
V LSD
700
600
to
800
4.1.14
100
V LSD
300
200
to
400
4.2
Alternating–Voltage Measurements
TI Range
Calibrator
(V ac)
(V ac)
(V ac)
4.2.2
2 V
400 Hz–1 V
1.000
0.993
to
1.007
”
20 V
400 Hz–10 V
10.00
9.93
to
10.07
”
20 V
4 kHz–10 V
10.00
9.93
to
10.07
”
20 V
50 kHz–10 V
10.00
9.47
to
10.53
”
200 V
50 kHz–10 V
10.00
9.47
to
10.53
”
200 V
4 kHz–10 V
10.00
9.93
to
10.07
”
200 V
400 Hz–100 V
100.0
99.3
to
100.7
”
750 V
400 Hz–100 V
100.0
99.3
to
100.7
NAVAIR 17–20VQ–05
153