TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Cubic 242 Instrument System
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 1
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+10 V
10 V x 0
0.000
–0.001
to
0.001
”
+10 V
10 V x .1
1.000
0.999
to
1.001
”
+10 V
10 V x .2
2.000
1.999
to
2.001
”
+10 V
10 V x .3
3.000
2.999
to
3.001
”
+10 V
10 V x .4
4.000
3.999
to
4.001
”
+10 V
10 V x .5
5.000
4.998
to
5.002
”
+10 V
10 V x .6
6.000
5.998
to
6.002
”
+10 V
10 V x .7
7.000
6.998
to
7.002
”
+10 V
10 V x .8
8.000
7.998
to
8.002
”
+10 V
10 V x .9
9.000
8.998
to
9.002
4.1.6
–10 V
10 V x 1
10.000
9.998
to
10.002
4.1.9
100 V
100
100.00
9.998
to
10.002
”
1000 V
100
100.0
99.9
to
100.1
Affix a special calibration label stating:
“The ratio measurement function has not been calibrated.”
NAVAIR 17–20VQ–05
(134 blank) 133