background image

 

TEST INST (S)

 

CALIBRATION CHECKLIST

 

PROC. NO.

Dana 3800, 3800A, 3860A, or 429883–01–01 Digital Voltmeter

 

MFG.

 

MODEL

 

SER.  NO.

PROCEDURE
    STEP
      NO.
      (1)

 

FUNCTION TESTED

            (2)

 

NOMINAL

      (3)

MEASURED VALUES

FIRST RUN

     (4)

SECOND RUN

        (5)

OUT
 OF
TOL
(6)

 

CALIBRATION TOLERANCES

                    (7)

NA 17–20VQ–05

Page   1   of   2

4.1

Direct–Voltage Measurements

TI Range

Calibrator

 (V)

(V)

(V)

4.1.3

+20 V

10 V x  0

0.00

–0.01

to

0.01

   ”

+20 V

10 V x .1

1.00

0.99

to

1.01

   ”

+20 V

10 V x .2

2.00

1.99

to

2.01

   ”

+20 V

10 V x .3

3.00

2.99

to

3.01

   ”

+20 V

10 V x .4

4.00

3.99

to

4.01

   ”

+20 V

10 V x .5

5.00

4.98

to

5.02

   ”

+20 V

10 V x .6

6.00

5.98

to

6.02

   ”

+20 V

10 V x .7

7.00

6.98

to

7.02

   ”

+20 V

10 V x .8

8.00

7.98

to

8.02

   ”

+20 V

10 V x .9

9.00

8.98

to

9.02

   ”

+20 V 

10 V x  1

10.00

9.98

to

10.02

4.1.6

–20 V

10 V x  1

10.00

9.98

to

10.02

TI Range Calibrator

mV)

(mV)

(mV)

4.1.9

200 mV (3860A) 1 V x .1

100.0

99.8

to

100.2

TI Range

Calibrator

 (V)

(V)

(V)

4.1.9

2 V

1 V x  1

1.000

0.998

to

1.002

   ”

200 V

100 V

100.0

99.8

to

100.2

   ”

2000 V

100 V

100

99

to

101

TI 200 mV Range (3860A)

V)

(

V)

(

V)

4.1.13

100 

V LSD

700

600

to

800

4.1.14

100 

V LSD

300

200

to

400

4.2

Alternating–Voltage Measurements 

Range

Calibrator

 (V)

(V)

(V)  

4.2.2

20 V

400 Hz–10 V

10.00

9.96

to

10.04

   ”

20 V

4 kHz–10 V

10.00

9.95

to

10.05

   ”

20 V

50 kHz–10 V

10.00

9.90

to

10.10 

   ” 

2 V

400 Hz–1 V

1.000

0.996

to

1.004

   ”

200 V

400 Hz–100 V

100.0

99.6

to

100.4

   ”

2000 V

400 Hz–100 V

100

99

to

101

NAVAIR 17–20VQ–05

85

Summary of Contents for 17-20VQ-05

Page 1: ...EIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONALPURPOSESONLY OTHERREQUESTSFORTHISDOCUMENTSHALLBEREFERREDTONAVY MEASUREMENT SCIENCE DIRECTORATE NAV...

Page 2: ...by a vertical line in the outer margin of the page Page No Revision Page No Revision Page No Revision Title 0 109 to 119 0 A 0 121 to 133 0 i 0 135 0 1 to 5 0 137 0 7 0 139 to 145 0 9 0 147 0 11 to 13...

Page 3: ...erations 3 4 Calibration Process 3 4 1 Direct Voltage Measurements 4 4 2 Alternating Voltage Measurements 5 4 3 Resistance Measurements 5 Navy Calibration Checklist 7 Appendix A Representative Test In...

Page 4: ...ated that the requirements for such calibrations will be similarly limited 1 4 All comments concerning this procedure should be directed to Navy Measurement Science Directorate Naval Warfare Assessmen...

Page 5: ...this procedure were selected from those known to be available at Navy calibration facilities and the listing by make or model number carries no implication of preference recommendation or approval for...

Page 6: ...the power switch is on it will automatically switch to battery power when unplugged from the ac line It may be warmed up in transit to another location or if previously warmed up it may be transported...

Page 7: ...controls for minimum output 4 1 8 Reverse the input leads to the TI for positive polarity 4 1 9 Set the calibrator V dc controls for a voltage value at the highest compatible calibration point for eac...

Page 8: ...itches for minimum output 4 3 RESISTANCE MEASUREMENTS 4 3 1 Set the TI controls to measure resistance 4 3 2 Press the calibrator ohms switch and set the calibrator output switch to 1 k 10 k 100 k 1000...

Page 9: ...ANCES 7 NA 17 20VQ 05 Page 1 of 1 4 1 Direct Voltage Measurements TI Range Calibrator V V V 4 1 3 10 V 10 V x 0 0 000 0 001 to 0 001 10 V x 1 1 000 0 999 to 1 001 10 V x 2 2 000 1 999 to 2 001 10 V x...

Page 10: ...Measurements TI Range Calibrator V V V 4 1 3 10 V 10 V x 0 0 000 0 001 to 0 001 10 V x 1 1 000 0 999 to 1 001 10 V x 2 2 000 1 999 to 2 001 10 V x 3 3 000 2 999 to 3 001 10 V x 4 4 000 3 999 to 4 001...

Page 11: ...to 7 0010 10 V x 8 8 0000 7 9989 to 8 0011 10 V x 9 9 0000 8 9988 to 9 0012 10 V x 1 10 0000 9 9987 to 10 0013 4 1 6 10 V 10 V x 1 10 0000 9 9987 to 10 0013 TI Range Calibrator mV mV mV 4 1 9 1000 mV...

Page 12: ...0 99800 to 1 00200 10 k range 10 0000 9 9800 to 10 0200 100 k range 100 000 99 800 to 100 200 1000 k range 1000 00 998 00 to1002 00 M M M 4 3 2 10 M range 10 0000 9 9800 to 10 0200 Affix a special cal...

Page 13: ...2 9992 to 3 0008 10 V 10 V x 4 4 0000 3 9990 to 4 0010 10 V 10 V x 5 5 0000 4 9987 to 5 0013 10 V 10 V x 6 6 0000 5 9985 to 6 0015 10 V 10 V x 7 7 0000 6 9982 to 7 0018 10 V 10 V x 8 8 0000 7 9980 to...

Page 14: ...to 2 0003 10 V 10 V x 3 3 0000 2 9997 to 3 0003 10 V 10 V x 4 4 0000 3 9996 to 4 0004 10 V 10 V x 5 5 0000 4 9996 to 5 0004 10 V 10 V x 6 6 0000 5 9996 to 6 0004 10 V 10 V x 7 7 0000 6 9995 to 7 0005...

Page 15: ...5 000 4 984 to 5 016 10 V 10 V x 6 6 000 5 982 to 6 018 10 V 10 V x 7 7 000 6 980 to 7 020 10 V 10 V x 8 8 000 7 979 to 8 021 10 V 10 V x 9 9 000 8 977 to 9 023 10 V 10 V x 1 10 000 9 975 to 10 025 10...

Page 16: ...0000 6 9987 to 7 0013 10 V 10 V x 8 8 0000 7 9986 to 8 0014 10 V 10 V x 9 9 0000 8 9984 to 9 0016 10 V 10 V x 1 10 0000 9 9983 to 10 0017 10 V 10 V x 1 10 0000 9 9983 to 10 0017 TI Range Calibrator V...

Page 17: ...calibration label stating 1 The 0 1 k range has not been calibrated 2 All other resistance ranges have been calibrated to 0 2 of range 3 The V ac function of the TI has been calibrated up to 50 kHz on...

Page 18: ...0 V 10 V x 0 0 000 0 001 to 0 001 10 V 10 V x 1 1 000 0 999 to 1 001 10 V 10 V x 2 2 000 1 999 to 2 001 10 V 10 V x 3 3 000 2 999 to 3 001 10 V 10 V x 4 4 000 3 999 to 4 001 10 V 10 V x 5 5 000 4 999...

Page 19: ...V V V 4 1 3 10 V 10 V x 0 0 00 0 01 to 0 01 10 V 10 V x 1 1 00 0 99 to 1 01 10 V 10 V x 2 2 00 1 99 to 2 01 10 V 10 V x 3 3 00 2 98 to 3 02 10 V 10 V x 4 4 00 3 98 to 4 02 10 V 10 V x 5 5 00 4 98 to 5...

Page 20: ...10 V x 6 6 000 5 997 to 6 003 10 V 10 V x 7 7 000 6 997 to 7 003 10 V 10 V x 8 8 000 7 997 to 8 003 10 V 10 V x 9 9 000 8 996 to 9 004 4 1 6 10 V 10 V x 9 9 000 9 996 to 9 004 TI Range Calibrator V V...

Page 21: ...easurements TI Calibrator k k k 4 3 1 k range 1 0000 9 9980 to 1 0020 10 k range 10 000 9 980 to 10 020 100 k range 100 00 99 80 to 100 20 1000 k range 1000 0 998 0 to 1002 0 10000 k range 10000 9980...

Page 22: ...2 2 000 1 999 to 2 001 10 V 10 V x 3 3 000 2 999 to 3 001 10 V 10 V x 4 4 000 3 999 to 4 001 10 V 10 V x 5 5 000 4 998 to 5 002 10 V 10 V x 6 6 000 5 998 to 6 002 10 V 10 V x 7 7 000 6 998 to 7 002 10...

Page 23: ...2 10 V 10 V x 7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 998 to 8 002 10 V 10 V x 9 9 000 8 998 to 9 002 10 V 10 V x 1 10 000 9 998 to 10 002 4 1 6 10 V 10 V x 1 10 000 9 998 to 10 002 4 1 9 1 V 1 V...

Page 24: ...Measurements with 225 or 230 converter TI Calibrator k k k 4 3 1 k range 1 0000 0 9980 to 1 0020 10 k range 10 000 9 980 to 10 020 100 k range 100 00 99 80 to 100 20 1000 k range 1000 0 998 0 to 1002...

Page 25: ...002 20 V 10 V x 6 6 000 5 998 to 6 002 20 V 10 V x 7 7 000 6 998 to 7 002 20 V 10 V x 8 8 000 7 998 to 8 002 20 V 10 V x 9 9 000 8 998 to 9 002 20 V 10 V x 1 10 000 9 998 to 10 002 4 1 6 20 V 10 V x 1...

Page 26: ...10 V 10 V x 7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 998 to 8 002 10 V 10 V x 9 9 000 8 998 to 9 002 10 V 10 V x 1 10 000 9 998 to 10 002 4 1 6 10 V 10 V x 1 10 000 9 998 to 10 002 4 1 9 100 V 10...

Page 27: ...Resistance Measurements 07 TI Calibrator k k k 4 3 2 1 k range 1 0000 0 9980 to 1 0020 10 k range 10 000 9 980 to 10 020 100 k range 100 00 99 80 to 100 20 1000 k range 1000 0 998 0 to 1002 0 1 k rang...

Page 28: ...6 6 000 5 998 to 6 002 10 V 10 V x 7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 998 to 8 002 10 V 10 V x 9 9 000 8 998 to 9 002 10 V 10 V x 1 10 000 9 998 to 10 002 4 1 6 10 V 10 V x 1 10 000 9 998 to...

Page 29: ...2 of 2 4 3 Resistance Measurements TI Calibrator k k k 4 3 2 1 k range 1 0000 0 9980 to 1 0020 10 k range 10 000 9 980 to 10 020 100 k range 100 00 99 80 to 100 20 1000 k range 1000 0 998 0 to 1002 0...

Page 30: ...V x 7 7 0000 6 9989 to 7 0011 10 V 10 V x 8 8 0000 7 9988 to 8 0012 10 V 10 V x 9 9 0000 8 9987 to 9 0013 10 V 10 V x 1 10 0000 9 9986 to 10 0014 4 1 6 10 V 10 V x 1 10 0000 9 9986 to 10 0014 TI Rang...

Page 31: ...2 4 3 Resistance Measurements 03 Assigned 0 2 TI Calibrator k k k 4 3 2 1 k range 1 00000 0 99800 to 1 00200 10 k range 10 0000 9 9800 to 10 0200 100 k range 100 000 99 800 to 100 200 1000 k range 100...

Page 32: ...0 V x 5 5 00 4 98 to 5 02 15 V 10 V x 6 6 00 5 98 to 6 02 15 V 10 V x 7 7 00 6 98 to 7 02 15 V 10 V x 8 8 00 7 98 to 8 02 15 V 10 V x 9 9 00 8 98 to 9 02 15 V 10 V x 1 10 00 9 98 to 10 02 4 1 6 15 V 1...

Page 33: ...996 to 5 004 10 V 10 V x 6 6 000 5 996 to 6 004 10 V 10 V x 7 7 000 6 995 to 7 005 10 V 10 V x 8 8 000 7 995 to 8 005 10 V 10 V x 9 9 000 8 994 to 9 006 10 V 10 V x 1 10 000 9 994 to 10 006 4 1 6 10 V...

Page 34: ...20VQ 05 Page 2 of 2 4 3 Resistance Measurements TI Range Calibrator k k k 4 3 2 1 k range 1 0000 0 9980 to 1 0020 10 k range 10 000 9 980 to 10 020 100 k range 100 00 99 80 to 100 20 1000 k range 1000...

Page 35: ...5 003 10 V 10 V x 6 6 000 5 997 to 6 003 10 V 10 V x 7 7 000 6 997 to 7 003 10 V 10 V x 8 8 000 7 996 to 8 004 10 V 10 V x 9 9 000 8 996 to 9 004 4 1 3 10 V 10 V x 1 10 000 9 996 to 10 004 4 1 6 10 V...

Page 36: ...LERANCES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements TI Calibrator k k k 4 3 2 1 k range 1 0000 0 9980 to 1 0020 10 k range 10 000 9 980 to 10 020 100 k range 100 00 99 80 to 100 20 1000 k...

Page 37: ...V 10 V x 6 6 000 5 996 to 6 004 10 V 10 V x 7 7 000 6 996 to 7 004 10 V 10 V x 8 8 000 7 995 to 8 005 10 V 10 V x 9 9 000 8 995 to 9 005 10 V 10 V x 1 10 000 9 995 to 10 005 4 1 6 10 V 10 V x 1 10 000...

Page 38: ...age 2 of 2 4 3 Resistance Measurements 03 TI Calibrator kW kW kW 4 3 2 1 kW 1 0000 0 9980 to 1 0020 10 kW 10 000 9 980 to 10 020 100 kW 100 00 99 80 to 100 20 1000 kW 1000 0 998 0 to 1002 0 MW MW MW 1...

Page 39: ...x 4 4 0000 3 9996 to 4 0004 10 V 10 V x 5 5 0000 4 9995 to 5 0005 10 V 10 V x 6 6 0000 5 9994 to 6 0006 10 V 10 V x 7 7 0000 6 9993 to 7 0007 10 V 10 V x 8 8 0000 7 9992 to 8 0008 10 V 10 V x 9 9 0000...

Page 40: ...ES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements TI Calibrator k k k 4 3 2 1 k 1 00000 0 99800 to 1 00200 10 k 10 0000 9 9800 to 10 0200 100 k 100 000 99 800 to 100 200 1000 k 1000 00 998 00...

Page 41: ...10 V 10 V x 6 6 0000 5 9993 to 6 0007 10 V 10 V x 7 7 0000 6 9992 to 7 0008 10 V 10 V x 8 8 0000 7 9991 to 8 0009 10 V 10 V x 9 9 0000 8 9990 to 9 0010 10 V 10 V x 1 10 0000 9 9989 to 10 0011 4 1 6 1...

Page 42: ...tance Measurements 02 Range Calibrator 4 3 2 1000 1000 00 998 00 to 1002 00 10000 10000 0 9980 0 to 10020 0 Range Calibrator k k k 4 3 2 100 k 100 000 99 800 to 100 200 1000 k 1000 00 998 00 to 1002 0...

Page 43: ...0 V 10 V x 5 5 000 4 998 to 5 002 10 V 10 V x 6 6 000 5 998 to 6 002 10 V 10 V x 7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 998 to 8 002 10 V 10 V x 9 9 000 8 998 to 9 002 4 1 6 10 V 10 V x 9 9 000...

Page 44: ...T RUN 4 SECOND RUN 5 OUT OF TOL 6 CALIBRATION TOLERANCES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements TI Calibrator k k k 4 3 2 1 k range 1 0000 0 9980 to 1 0020 10 k range 10 000 9 980 to...

Page 45: ...V x 7 7 00 6 98 to 7 02 10 V 10 V x 8 8 00 7 98 to 8 02 10 V 10 V x 9 9 00 8 98 to 9 02 10 V 10 V x 1 10 00 9 98 to 10 02 4 1 6 10 V 10 V x 1 10 00 9 98 to 10 02 TI Range Calibrator mV mV mV 4 1 9 100...

Page 46: ...BRATION TOLERANCES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements TI Calibrator k k k 4 3 2 1 k range 1 000 0 998 to 1 002 10 k range 10 00 9 98 to 10 02 100 k range 100 0 99 8 to 100 2 M M M...

Page 47: ...10 V 10 V x 7 7 000 6 996 to 7 004 10 V 10 V x 8 8 000 7 996 to 8 004 10 V 10 V x 9 9 000 8 996 to 9 004 10 V 10 V x 1 10 000 9 996 to 10 004 4 1 6 10 V 10 V x 1 10 000 9 996 to 10 004 TI Range Calib...

Page 48: ...2 1000 1000 0 998 0 to 1002 0 Range Calibrator k k k 4 3 2 10 k 10 000 9 980 to 10 020 100 k 100 00 99 80 to 100 20 1000 k 1000 0 998 0 to 1002 0 M M M 4 3 2 10 M 10 000 9 980 to 10 020 Affix a specia...

Page 49: ...000 6 9984 to 7 0016 10 V 10 V x 8 8 0000 7 9983 to 8 0017 10 V 10 V x 9 9 0000 8 9981 to 9 0019 10 V 10 V x 1 10 0000 9 9980 to 10 0020 4 1 6 10 V 10 V x 1 10 0000 9 9980 to 10 0020 TI Range Calibrat...

Page 50: ...Measurements Assigned 0 2 TI Range Calibrator k k k 4 3 2 1 k 1 00000 0 99800 to 1 00200 10 k 10 0000 9 9800 to 10 0200 100 k 100 000 99 800 to 100 200 1000 k 1000 00 998 00 to 1002 00 M M m 4 3 2 10...

Page 51: ...000 6 9991 to 7 0009 10 V 10 V x 8 8 0000 7 9990 to 8 0010 10 V 10 V x 9 9 0000 8 9989 to 9 0011 10 V 10 V x 1 10 0000 9 9988 to 10 0012 4 1 6 10 V 10 V x 1 10 0000 9 9988 to 10 0012 TI Range Calibrat...

Page 52: ...595 1000 V 400 Hz 100 V 100 00 99 47 to 100 53 1000 V 4 kHz 10 V 10 00 9 59 to 10 41 4 3 Resistance Measurements TI Range Calibrator 4 3 2 1000 1000 00 998 00 to 1002 00 10000 10000 0 99800 to 10020 0...

Page 53: ...3 10 V 10 V x 0 0 00 0 01 to 0 01 10 V 10 V x 1 1 00 0 99 to 1 01 10 V 10 V x 2 2 00 1 99 to 2 01 10 V 10 V x 3 3 00 2 99 to 3 01 10 V 10 V x 4 4 00 3 99 to 4 01 10 V 10 V x 5 5 00 4 98 to 5 02 10 V...

Page 54: ...8 8 0000 7 9983 to 8 0017 10 V 10 V x 9 9 0000 8 9981 to 9 0019 10 V 10 V x 1 10 0000 9 9980 to 10 0020 4 1 6 10 V 10 V x 1 10 0000 0 9980 to 10 0020 TI Range Calibrator mV mV mV 4 1 9 1000 mV 1 x 1 1...

Page 55: ...4 3 Resistance Measurements 8300A 02 Assigned 0 2 TI Calibrator k k k 4 3 2 1 k range 1 00000 0 99800 to 1 00200 10 k range 10 0000 9 9800 to 10 0200 100 k range 100 000 99 800 to 100 200 1000 k range...

Page 56: ...000 7 990 to 8 010 20 V 10 V x 9 9 000 8 989 to 9 011 20 V 10 V x 1 10 000 9 988 to 10 012 4 1 6 20 V 10 V x 1 10 000 9 998 to 10 012 TI Range Calibrator V V V 4 1 9 2 V 1 V x 1 1 0000 0 9988 to 1 001...

Page 57: ...esistance Measurements TI Calibrator kW kW kW 4 3 2 2 kW 1 kW 1 0000 0 9968 to 1 0032 20 kW 10 kW 10 000 9 968 to 10 032 200 kW 100 kW 100 00 99 68 to 100 32 TI Calibrator MW MW MW 4 3 2 2 MW 1 MW 1 0...

Page 58: ...7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 998 to 8 002 10 V 10 V x 9 9 000 8 998 to 9 002 4 1 6 10 V 10 V x 1 10 000 9 997 to 10 003 TI Range Calibrator V V V 4 1 9 1 V 1 V x 1 1 0000 0 9997 to 1 0...

Page 59: ...Range Calibrator k k k 4 3 2 1 k 1 k 1 0000 0 9980 to 1 0020 10 k 10 k 10 000 9 994 to 10 006 100 k 100 k 100 00 99 94 to 100 06 1000 k 1 M 1000 0 998 9 to 1001 1 10000 k 10 M 10000 9974 to 10026 Aff...

Page 60: ...6 9993 to 7 0007 10 V 10 V x 8 8 0000 7 9992 to 8 0008 10 V 10 V x 9 9 0000 8 9991 to 9 0009 4 1 6 10 V 10 V x 1 10 0000 0 9990 to 10 0010 TI Range Calibrator V V V 4 1 9 1 V 1 V x 1 1 00000 0 99990...

Page 61: ...k 4 3 2 1 k 1 k 1 00000 0 99847 to 1 00153 10 k 10 k 10 0000 9 9847 to 10 0153 100 k 100 k 100 000 99 847 to 100 153 1000 k 1000 k 1000 00 998 47 to 1001 53 10000 k 10000 k 10000 0 9980 0 to 10020 0...

Page 62: ...0 V x 7 7 0000 6 9992 to 7 0008 10 V 10 V x 8 8 0000 7 9991 to 8 0009 10 V 10 V x 9 9 0000 8 9990 to 9 0010 4 1 6 10 V 10 V x 9 9 0000 8 9990 to 9 0010 TI Range Calibrator mV mV mV 4 1 9 1000 mV 1 V x...

Page 63: ...1 0000 0 9980 to 1 0020 10 k 10 k 10 0000 ck 9 9800 to 10 0200 100 k 10 k 10 000 9 980 to 10 020 100 k 100 k 100 000 ck 99 800 to 100 200 TI Range Calibrator M M M 4 3 2 1 M 100 k 0 100000 0 09980 to...

Page 64: ...7 7 00 6 98 to 7 02 10 V 10 V x 8 8 00 7 98 to 8 02 10 V 10 V x 9 9 00 8 98 to 9 02 10 V 10 V x 1 10 00 9 98 to 10 02 4 1 6 10 V 10 V x 1 10 00 9 98 to 10 02 TI Range Calibrator mV mV mV 4 1 9 100 mV...

Page 65: ...100 V 100 99 to 101 4 3 Resistance Measurements TI Range Calibrator 4 3 2 100 100 0 98 7 to 101 3 TI Range Calibrator k k k 1 k 1 000 0 987 to 1 013 10 k 10 00 9 87 to 10 13 100 k 100 0 98 7 to 101 3...

Page 66: ...0008 10 V 10 V x 9 9 0000 8 9991 to 9 0009 4 1 6 10 V 10 V x 9 9 0000 8 9991 to 9 0009 TI Range Calibrator V V V 4 1 9 100 V 10 V x 1 10 000 9 9990 to 10 001 100 V 100 V 100 000 ck 99 990 to 100 01 1...

Page 67: ...0011 10 V 10 V x 2 2 000 1 998 to 2 001 10 V 10 V x 3 3 000 2 999 to 3 001 10 V 10 V x 4 4 000 3 999 to 4 001 10 V 10 V x 5 5 000 4 998 to 5 002 10 V 10 V x 6 6 000 5 998 to 6 002 10 V 10 V x 7 7 000...

Page 68: ...4 10 V 10 V x 7 7 000 6 995 to 7 005 10 V 10 V x 8 8 000 7 995 to 8 005 10 V 10 V x 9 9 000 8 994 to 9 006 4 1 6 10 V 10 V x 1 10 000 9 994 to 10 006 4 1 9 TI Range Calibrator mV mV mV 100 mV 1 x 1 10...

Page 69: ...librator V V V 10 V 50 kHz 10 V 10 000 9 978 to 10 022 4 3 Resistance Measurements k Ranges k k 4 3 2 1 1 0000 0 9969 to 1 0031 10 10 000 9 969 to 10 031 100 100 00 99 69 to 100 31 1000 1000 0 996 9 t...

Page 70: ...to 7 003 10 V 10 V x 8 8 000 7 997 to 8 003 10 V 10 V x 9 9 000 8 996 to 9 004 10 V 10 V x 1 10 000 9 996 to 10 004 4 1 6 10 V 10 V x 1 10 000 9 996 to 10 004 TI Range Calibrator V V V 4 1 9 1 V 1 V x...

Page 71: ...3 Resistance Measurements TI Calibrator k k k 4 3 2 1 k 1 00000 0 9980 to 1 0020 10 k 10 0000 9 980 to 10 020 100 k 100 000 99 80 to 100 20 1000 k 1000 00 998 0 to 1002 0 10000 k 1000 00 9980 to 10020...

Page 72: ...00 6 98 to 7 02 10 V 10 V x 8 8 00 7 98 to 8 02 10 V 10 V x 9 9 00 8 98 to 9 02 10 V 10 V x 1 10 00 9 98 to 10 02 4 1 6 10 V 10 V x 1 10 00 9 98 to 10 02 TI Range Calibrator mV mV mV 4 1 9 100 mV 1 V...

Page 73: ...V 100 97 to 103 4 3 Resistance Measurements TI Range Calibrator 4 3 2 100 100 0 98 6 to 101 4 TI Range Calibrator k k k 4 3 2 1 k 1 k 1 000 0 986 to 1 014 10 k 10 k 10 00 9 86 to 10 14 100 k 100 k 10...

Page 74: ...0 V x 7 7 00 6 98 to 7 02 20 V 10 V x 8 8 00 7 98 to 8 02 20 V 10 V x 9 9 00 8 98 to 9 02 20 V 10 V x 1 10 00 9 98 to 10 02 4 1 6 20 V 10 V x 1 10 00 9 98 to 10 02 TI Range Calibrator mV mV mV 4 1 9 2...

Page 75: ...Range Calibrator k k k 4 3 2 2 k 1 k 1 000 0 997 to 1 003 20 k 10 k 10 00 9 97 to 10 03 200 k 100 k 100 0 99 7 to 100 3 TI Range Calibrator M M M 4 3 2 2 M 1 M 1 000 0 995 to 1 005 20 M 10 M 10 00 9 8...

Page 76: ...0 V 10 V x 6 6 0000 5 9995 to 6 0005 10 V 10 V x 7 7 0000 6 9994 to 7 0006 10 V 10 V x 8 8 0000 7 9993 to 8 0007 10 V 10 V x 9 9 0000 8 9993 to 9 0007 10 V 10 V x 1 10 0000 9 9992 to 10 0008 4 1 6 10...

Page 77: ...20VQ 05 Page 2 of 2 4 3 Resistance Measurements k Ranges k k k 4 3 2 1 1 00000 0 99800 to 1 00200 10 10 0000 9 9800 to 10 0200 100 100 000 99 800 to 100 200 1000 1000 00 998 00 to 1002 00 10 000 1000...

Page 78: ...x 5 5 0000 4 9994 to 5 0006 10 V 10 V x 6 6 0000 5 9993 to 6 0007 10 V 10 V x 7 7 0000 6 9992 to 7 0008 10 V 10 V x 8 8 0000 7 9991 to 8 0009 10 V 10 V x 9 9 0000 8 9990 to 9 0010 4 1 6 10 V 10 V x 9...

Page 79: ...truments stating the reasons as indicated as applicable 6801A DC Preamplifier The X100 voltage range has not been calibrated 6802A DC Preamplifier The X1000 and X100 voltage ranges have not been calib...

Page 80: ...V 10 V x 7 7 0000 6 9992 to 7 0008 10 V 10 V x 8 8 0000 7 9991 to 8 0009 10 V 10 V x 9 9 0000 8 9990 to 9 0010 4 1 6 10 V 10 V x 9 9 0000 8 9990 to 9 0010 TI Range Calibrator mV mV mV 4 1 9 1000 mV 1...

Page 81: ...M range 10 000 ck 9 9800 to 10 020 Affix special calibration labels to the TI as follows 1 If the TI has a ratio measurement function affix a special calibration label stating The ratio function has n...

Page 82: ...8 to 5 002 10 V 10 V x 6 6 000 5 998 to 6 002 10 V 10 V x 7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 997 to 8 003 10 V 10 V x 9 9 000 8 997 to 9 003 4 1 6 10 V 10 V x 1 10 000 9 997 to 10 003 TI Ran...

Page 83: ...nts k Ranges k k k 4 3 2 1 k 1 0000 0 9980 to 1 0020 10 k 10 000 9 994 to 10 006 100 k 100 00 99 94 to 100 06 M M M 4 3 2 10 M Range 10 000 9 980 to 10 020 Affix a special calibration label to the fol...

Page 84: ...0 8 9990 to 9 0010 4 1 6 10 V 10 V x 9 9 0000 8 9990 to 9 0010 TI Range Calibrator mV mV mV 4 1 9 1000 mV 1 x 0 9 900 00 899 82 to 900 18 TI Range Calibrator V V V 4 1 9 100 V 10 V 10 000 99 998 to 10...

Page 85: ...10 V x 6 6 0000 5 9988 to 6 0012 10 V 10 V x 7 7 0000 6 9987 to 7 0013 10 V 10 V x 8 8 0000 7 9986 to 8 0014 10 V 10 V x 9 9 0000 8 9985 to 9 0015 10 V 10 V x 1 10 0000 9 9984 to 10 0016 4 1 6 10 V 10...

Page 86: ...79 to 7 0021 20 V 10 V x 8 8 0000 7 9978 to 8 0022 20 V 10 V x 9 9 0000 8 9976 to 9 0024 4 1 6 20 V 10 V x 1 10 0000 9 9975 to 10 0025 4 1 9 2 V 1 V x 1 1 0000 0 99975 to 1 00025 200 V 100 V 100 000 9...

Page 87: ...10 kW 10 0000 9 9800 to 10 0200 200 kW 100 kW 100 000 99 800 to 100 200 2000 kW 1 MW 1000 00 998 00 to 1002 00 TI Range Calibrator MW MW MW 4 3 2 20 MW 10 MW 10 0000 9 9700 to 10 0300 W W W 200 W 100...

Page 88: ...0 V x 7 7 000 6 998 to 7 002 10 V 10 V x 8 8 000 7 997 to 8 003 10 V 10 V x 9 9 000 8 997 to 9 003 4 1 6 10 V 10 V x 1 10 000 9 997 to 10 003 4 1 9 1 V 1 V x 1 1 0000 0 9997 to 1 0003 100 V 100 V 100...

Page 89: ...k 4 3 2 1 k 1 0000 0 9980 to 1 0020 10 k 10 000 9 980 to 10 020 100 k 100 00 99 80 to 100 20 1000 k 1000 0 998 0 to 1002 0 TI Range Calibrator M M M 4 3 2 10 M range 10 000 9 980 to 10 020 Affix a spe...

Page 90: ...to 5 0007 10 V 10 V x 6 6 0000 5 9993 to 6 0007 10 V 10 V x 7 7 0000 6 9992 to 7 0008 10 V 10 V x 8 8 0000 7 9992 to 8 0008 10 V 10 V x 9 9 0000 8 9991 to 9 0009 4 1 6 10 V 10 V x 1 10 0000 9 9991 to...

Page 91: ...99 to 4 01 10 V 10 V x 5 5 00 4 98 to 5 02 10 V 10 V x 6 6 00 5 98 to 6 02 10 V 10 V x 7 7 00 6 98 to 7 02 10 V 10 V x 8 8 00 7 98 to 8 02 10 V 10 V x 9 9 00 8 98 to 9 02 4 1 6 10 V 10 V x 1 10 00 9...

Page 92: ...006 10 V 10 V x 8 8 000 7 994 to 8 006 10 V 10 V x 9 9 000 8 993 to 9 007 4 1 6 10 V 10 V x 1 10 000 9 993 to 10 007 4 1 9 1 V 1 V x 1 1 0000 0 9995 to 1 0005 100 V 100 V 100 00 99 93 to 100 07 1000...

Page 93: ...6 004 10 V 10 V x 7 7 000 6 995 to 7 005 10 V 10 V x 8 8 000 7 995 to 8 005 10 V 10 V x 9 9 000 8 994 to 9 006 4 1 6 10 V 10 V x 1 10 000 9 994 to 10 006 4 1 9 1 V 1 V x 1 1 00000 0 9994 to 1 0006 100...

Page 94: ...ance Measurements TI Range Calibrator k k k 4 3 2 1 k 1 k 1 0000 0 9980 to 1 0020 10 k 10 k 10 000 9 992 to 10 008 100 k 100 k 100 00 99 92 to 100 08 1000 k 1 M 1000 0 998 9 to 1001 1 TI Range Calibra...

Page 95: ...6 006 10 V 10 V x 7 7 000 6 993 to 7 007 10 V 10 V x 8 8 000 7 992 to 8 008 10 V 10 V x 9 9 000 8 992 to 9 008 10 V 10 V x 1 10 000 9 991 to 10 009 4 1 6 100 V 10 V x 1 10 000 9 991 to 10 009 TI Rang...

Page 96: ...OUT OF TOL 6 CALIBRATION TOLERANCES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements k Ranges k k k 4 3 2 1 1 0000 0 9968 to 1 0032 10 10 000 9 968 to 10 032 100 100 00 99 68 to 100 32 M Range...

Page 97: ...10 V 10 V x 7 7 000 6 997 to 7 003 10 V 10 V x 8 8 000 7 997 to 8 003 10 V 10 V x 9 9 000 8 997 to 9 003 10 V 10 V x 1 10 000 9 997 to 10 003 4 1 6 10 V 10 V x 1 10 000 9 997 to 10 003 TI Range Calibr...

Page 98: ...Resistance Measurements kW Ranges kW kW kW 4 3 2 1 1 0000 0 9980 to 1 0020 10 10 000 9 980 to 10 020 100 100 00 99 80 to 100 20 MW Ranges MW MW MW 4 3 2 1 1 0000 0 991 to 1 0009 10 10 000 9 998 to 10...

Page 99: ...to 9 0030 4 1 6 10 V 10 V x 9 9 0000 8 9970 to 9 0030 TI Range Calibrator V V V 4 1 9 100 V 10 x 1 10 000 9 994 to 10 006 100 V 100 100 000 ck 99 967 to 100 033 1000 V 100 100 00 99 60 to 100 40 TI 10...

Page 100: ...4 3 Resistance Measurements TI Range Calibrator k k k 4 3 2 10 k 1 k 1 0000 0 9947 to 1 0053 10 k 10 k 10 0000 ck 9 9497 to 10 0503 100 k 10 k 10 000 9 947 to 10 053 100 k 100 k 100 000 ck 99 497 to 1...

Page 101: ...0 V 10 V x 6 6 00 5 96 to 6 04 20 V 10 V x 7 7 00 6 96 to 7 04 20 V 10 V x 8 8 00 7 96 to 8 04 20 V 10 V x 9 9 00 8 96 to 9 04 4 1 6 20 V 10 V x 1 10 00 9 95 to 10 05 4 1 9 2 V 10 V x 1 1 000 0 995 to...

Page 102: ...TOL 6 CALIBRATION TOLERANCES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements TI Range Calibrator 4 3 2 200 100 k 1 000 0 997 to 1 003 TI Range Calibrator k k k 4 3 2 2 k 1 k 1 000 0 993 to 1...

Page 103: ...00 2 1000 mV 1 V x 2 200 0 199 8 to 200 2 1000 mV 1 V x 3 300 0 299 7 to 300 3 1000 mV 1 V x 4 400 0 399 7 to 400 3 1000 mV 1 V x 5 500 0 499 6 to 500 4 1000 mV 1 V x 6 600 0 599 6 to 600 4 1000 mV 1...

Page 104: ...0 V x 9 9 0000 8 9989 to 9 0011 4 1 6 10 V 10 V x 1 10 0000 9 9988 to 10 0012 4 1 9 100 mV 1 V x 1 10 0000 0 099910 to 0 100090 1 V 1 V x 1 1 00000 0 99988 to 1 00012 100 V 100 V 100 000 99 988 to 100...

Page 105: ...0 to 100 200 1000 k 1 M 1000 00 998 00 to 1002 00 10000 k 10 M 10000 0 9980 0 to 10020 0 Affix a special calibration label stating the following as applicable 1 The 100 mV of the direct voltage range...

Page 106: ...to 7 0009 10 V 10 V x 8 8 0000 7 9990 to 8 0010 10 V 10 V x 9 9 0000 8 9989 to 9 0011 4 1 6 10 V 10 V x 1 10 0000 9 9988 to 10 0012 4 1 9 1 V 1 V x 1 1 00000 0 99985 to 1 00015 4 1 9 100 V 100 V 100...

Page 107: ...0 99 800 to 100 200 1000 k 1 M 1000 00 998 00 to 1002 00 10000 k 10 M 10000 0 9980 0 to 10020 0 Affix a special calibration label stating the following information as applicable 1 The TI direct voltag...

Page 108: ...02 20 V 10 V x 7 7 00 6 98 to 7 02 20 V 10 V x 8 8 00 7 98 to 8 02 20 V 10 V x 9 9 00 8 98 to 9 02 4 1 6 20 V 10 V x 1 10 00 9 98 to 10 02 4 1 9 2 V 1 V x 1 1 000 0 998 to 1 002 200 V 100 V 100 0 99...

Page 109: ...ange Calibrator W W W 4 3 2 200 W 100 W 100 0 9943 to 100 6 TI Range Calibrator kW kW kW 4 3 2 2 kW 1 kW 1 000 0 997 to 1 003 20 kW 10 kW 10 00 9 97 to 10 03 200 kW 100 kW 100 0 99 7 to 100 3 2000 kW...

Page 110: ...V x 2 2 00 1 97 to 2 03 V 10 V x 3 3 00 2 96 to 3 04 V 10 V x 4 4 00 3 95 to 4 05 V 10 V x 5 5 00 4 94 to 5 06 V 10 V x 6 6 00 5 94 to 6 06 V 10 V x 7 7 00 6 93 to 7 07 V 10 V x 8 8 00 7 92 to 8 08 V...

Page 111: ...6 CALIBRATION TOLERANCES 7 NA 17 20VQ 05 Page 2 of 2 4 3 Resistance Measurements FUNCTION Calibrator k k k 4 3 2 k 1 k 1 000 0 983 to 1 017 k 10 k 10 00 9 83 to 10 17 k 100 k 100 0 98 3 to 101 7 k 1 M...

Page 112: ...20 V 10 V x 8 8 000 7 995 to 8 005 20 V 10 V x 9 9 000 8 994 to 9 006 4 1 6 20 V 10 V x 1 10 000 9 994 to 10 006 4 1 9 2 V 1 V x 1 1 0000 0 9994 to 1 0006 200 V 100 V 100 00 99 94 to 100 06 1200 V 100...

Page 113: ...99 36 to 100 64 TI Range Calibrator kW kW kW 2 kW 1 kW 1 0000 0 9968 to 1 032 20 kW 10 kW 10 000 9 968 to 10 032 200 kW 100 kW 100 00 99 68 to 100 32 2000 kW 1 MW 1000 0 996 8 to 1003 2 TI Range Cali...

Page 114: ...9965 to 8 00035 20 V 10 V x 9 9 0000 8 9961 to 9 0039 4 1 6 20 V 10 V x 1 10 0000 9 9957 to 10 0043 4 1 9 2 V 1 V x 1 1 00000 0 99957 to 1 00043 200 V 100 V 100 000 99 957 to 100 043 1200 V 100 V 100...

Page 115: ...200 20 k 10 k 10 0000 9 9800 to 10 0200 200 k 100 k 100 000 99 800 to 100 200 2000 k 1 M 1000 00 998 96 to 1001 04 TI Range Calibrator M M M 4 3 2 20 M 10 M 10 0000 9 9594 to 10 0406 Affix a special c...

Page 116: ...x 1 1 000 0 999 to 1 001 10 V 10 V x 2 2 000 1 999 to 2 001 10 V 10 V x 3 3 000 2 999 to 3 001 10 V 10 V x 4 4 000 3 999 to 4 001 10 V 10 V x 5 5 000 4 998 to 5 002 10 V 10 V x 6 6 000 5 998 to 6 002...

Page 117: ...V 10 V x 9 9 00 8 92 to 9 08 4 1 6 20 V 10 V x 1 10 00 9 91 to 10 09 4 1 9 2 V 1 V x 1 1 000 0 994 to 1 006 100 V 100 V 100 0 99 1to 100 9 1 kV 100 V 100 99to 101 4 2 Alternating Voltage Measurements...

Page 118: ...00 1 9998 to 2 0002 Auto 10 V x 3 3 0000 2 9997 to 3 0003 Auto 10 V x 4 4 0000 3 9996 to 4 0004 Auto 10 V x 5 5 0000 4 9995 to 5 0005 Auto 10 V x 6 6 0000 5 9994 to 6 0006 Auto 10 V x 7 7 0000 6 9993...

Page 119: ...to 7 02 20 V 10 V x 8 8 00 7 98 to 8 02 20 V 10 V x 9 9 00 8 98 to 9 02 4 1 6 20 V 10 V x 1 10 00 9 98 to 10 02 TI Range Calibrator mV mV mV 4 1 9 200 mV 1 V x 0 1 100 0 99 8 to 100 2 TI Range Calibra...

Page 120: ...100 9 TI Range Calibrator kW kW kW 4 3 2 2 kW 1 kW 1 000 0 940 to 1 006 20 kW 10 kW 10 00 9 94 to 10 06 200 kW 100 kW 100 0 99 4 to 100 6 TI Range Calibrator MW MW MW 4 3 2 2 MW 1 MW 1 000 0 994 to 1...

Page 121: ...002 10 x 6 6 000 5 998 to 6 002 10 x 7 7 000 6 998 to 7 002 10 x 8 8 000 7 998 to 8 002 10 x 9 9 000 8 998 to 9 002 10 x 1 10 000 9 998 to 10 002 4 1 6 20 V 10 x 1 10 000 9 998 to 10 002 4 1 9 0 2 V 1...

Page 122: ...se has been calibrated only at 400 Hz 4 kHz and 50 kHz At 50 kHz the test instrument is calibrated to 0 14 iv 16 digits 3 The AC DC current functions have not been calibrated 4 The 200 range has been...

Page 123: ...V 10 V x 6 6 00 5 91 to 6 09 10 V 10 V x 7 7 00 6 90 to 7 10 10 V 10 V x 8 8 00 7 90 to 8 10 10 V 10 V x 9 9 00 8 89 to 9 11 4 1 6 10 V 10 V x 9 9 00 8 89 to 9 11 4 1 9 1000 mV 1 x 9 0 900 0 889 to 0...

Page 124: ...ATION TOLERANCES 7 NA 17 20VQ 05 Page 2 of 2 Range Calibrator M M M 4 3 2 10 M 1 M 1 00 0 87 to 1 13 10 M 10 M 10 0 9 74 to 10 26 Affix a special calibration label on the test instrument stating The V...

Page 125: ...7 02 20 V 10 x 8 8 00 7 98 to 8 02 20 V 10 x 9 9 00 8 98 to 9 02 20 V 10 x 1 10 00 9 98 to 10 02 4 1 6 20 V 10 x 1 10 00 9 98 to 10 02 4 1 9 0 2 V 1 x 1 0 100 0 0998 to 0 1002 2 V 10 x 1 1 000 0 998...

Page 126: ...000 5 9988 to 6 0012 10 V 10 V x 7 7 0000 6 9987 to 7 0013 10 V 10 V x 8 8 0000 7 9986 to 8 0014 10 V 10 V x 9 9 0000 8 9985 to 9 0015 10 V 10 V x 1 10 0000 9 9984 to 10 0016 4 1 6 10 V 10 V x 1 1 000...

Page 127: ...02 20 V 10 V x 8 8 00 7 98 to 8 02 20 V 10 V x 9 9 00 8 98 to 9 02 4 1 6 20 V 10 V x 1 10 00 9 98 to 10 02 4 1 9 2 V 1 V x 1 1 000 0 998 to 1 002 200 V 100 V 100 0 99 8 to 100 2 1200 V 100 V 100 99 to...

Page 128: ...TI Range Calibrator W W W 4 3 2 200 W 100 W 100 0 99 4 to 100 6 TI Range Calibrator kW kW kW 4 3 2 2 kW 1 kW 1 000 0 997 to 1 003 20 kW 10 kW 10 00 9 97 to 10 03 200 kW 100 kW 100 0 99 7 to 100 3 2000...

Page 129: ...10 V 10 V x 6 6 00000 5 9991 0 to 6 0009 0 10 V 10 V x 7 7 00000 6 9991 0 to 7 0009 0 10 V 10 V x 8 8 00000 7 9991 0 to 8 0009 0 10 V 10 V x 9 9 00000 8 9991 0 to 9 0009 0 10 V 10 V x 1 10 00000 9 999...

Page 130: ...3 Resistance Measurements 5789 TI Range Calibrator W W W 4 3 2 1 kW kW range 1000 0 998 000 to 1002 00 10 kW kW range 10000 0 9980 0 to 10020 0 100 kW kW range 100000 99980 0 to 100020 TI Range Calib...

Page 131: ...1 6 20 V 10 x 9 9 00 8 98 to 9 02 TI Range Calibrator V dc V dc V dc 4 1 9 200 mV 1 V x 0 1 100 0 99 8 to 100 2 mV 2 V 1 V x 1 1 000 0 998 to 1 002 200 V 100 V 100 0 99 8 to 100 2 1000 V 100 V 100 0 9...

Page 132: ...3 2 200 W 1 W 0 1000 0 0997 to 1003 4 3 2 2 kW 1 kW 1 000 0 997 to 1 003 20 kW 10 kW 10 00 9 97 to 10 03 200 kW 100 0 kW 100 0 99 7 to 100 3 TI Range Calibrator MW MW MW 4 3 2 2000 MW 1 000 1 000 0 9...

Page 133: ...o 7 03 20 V 10 x 9 9 00 8 96 to 9 04 4 1 6 20 V 10 x 9 9 00 8 96 to 9 04 TI Range Calibrator V dc V dc V dc 4 1 9 200 mV 1 V x 1 100 0 mV 99 6to 100 4 mV 2 V 1 V x 1 1 000 0 996to 1 004 200 V 100 V 10...

Page 134: ...surements 4 3 2 TI Range Calibrator k k k 2 k 1 k 1 000 0 940 to 1 006 20 k 10 k 10 00 9 94 to 10 06 200 k 100 k 100 0 99 4 to 100 6 TI Range Calibrator M M M 4 3 2 2 k 1 M 1 000 0 994 to 1 006 20 M 1...

Page 135: ...to 202 1 320 mV 1 V x 3 300 0 296 9 to 303 1 TI Range Calibrator V V V 4 1 9 32 V 10 V x 1 10 00 9 89 to 10 11 320 V 100 V 100 0 98 9 to 101 1 1000 V 100 V 100 97 7 to 102 3 TI 320 mV Range V V V 4 1...

Page 136: ...nd AC current functions have not been calibrated 2 The 32 and 320 V ranges both DC and AC function have been calibrated only at approx 1 3 of full range 3 The TI 1000 V dc and 750 V ac ranges have bee...

Page 137: ...47 ME 346 ASM 308 Digital Voltmeter 71 TECH310 Beckman Instruments Digital Multimeter 155 V35 Non Linear System Digital Voltmeter 75 V35A V35B V35BR V72 1000 1 Cubic 7 0309020 9 0309020 1 111 Simpson...

Page 138: ...3446A AC DC Remote Unit 3460B Digital Voltmeter 15 3465A 101 34703AOPTE01 Marden System 17 34740A 34703A Digital Multimeter 107 3490A 19 350 Data Tech Corp 81 3500A Electro Instruments DC Voltmeter R...

Page 139: ...24 6770 AC Converter 91 6771 6801A DC Preamplifier 89 6802A 6980B AC Converter 71 95 6990 Ohm Converter 91 6910A 89 6911A 7000A Digital Multimeter 7005 Systron Donner Digital Volt Multimeter 37 7005A...

Page 140: ...001 8400AOPT002 8400AOPT005 8400AOPT007 8425A 8425A AF 8425B 8425BOPT001 8425BOPT002 8500A AEOPT1 2 5 8 151 851 8014A Electro Instruments Digital Multimeter 51 851 8014B 8600A Fluke 129 8600AOPT001 86...