TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Dana 3800, 3800A, 3860A, or 429883–01–01 Digital Voltmeter
MFG.
MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20VQ–05
Page 1 of 2
4.1
Direct–Voltage Measurements
TI Range
Calibrator
(V)
(V)
(V)
4.1.3
+20 V
10 V x 0
0.00
–0.01
to
0.01
”
+20 V
10 V x .1
1.00
0.99
to
1.01
”
+20 V
10 V x .2
2.00
1.99
to
2.01
”
+20 V
10 V x .3
3.00
2.99
to
3.01
”
+20 V
10 V x .4
4.00
3.99
to
4.01
”
+20 V
10 V x .5
5.00
4.98
to
5.02
”
+20 V
10 V x .6
6.00
5.98
to
6.02
”
+20 V
10 V x .7
7.00
6.98
to
7.02
”
+20 V
10 V x .8
8.00
7.98
to
8.02
”
+20 V
10 V x .9
9.00
8.98
to
9.02
”
+20 V
10 V x 1
10.00
9.98
to
10.02
4.1.6
–20 V
10 V x 1
10.00
9.98
to
10.02
TI Range Calibrator
mV)
(mV)
(mV)
4.1.9
200 mV (3860A) 1 V x .1
100.0
99.8
to
100.2
TI Range
Calibrator
(V)
(V)
(V)
4.1.9
2 V
1 V x 1
1.000
0.998
to
1.002
”
200 V
100 V
100.0
99.8
to
100.2
”
2000 V
100 V
100
99
to
101
TI 200 mV Range (3860A)
V)
(
V)
(
V)
4.1.13
100
V LSD
700
600
to
800
4.1.14
100
V LSD
300
200
to
400
4.2
Alternating–Voltage Measurements
Range
Calibrator
(V)
(V)
(V)
4.2.2
20 V
400 Hz–10 V
10.00
9.96
to
10.04
”
20 V
4 kHz–10 V
10.00
9.95
to
10.05
”
20 V
50 kHz–10 V
10.00
9.90
to
10.10
”
2 V
400 Hz–1 V
1.000
0.996
to
1.004
”
200 V
400 Hz–100 V
100.0
99.6
to
100.4
”
2000 V
400 Hz–100 V
100
99
to
101
NAVAIR 17–20VQ–05
85