A
DJUSTING
THE
MEASUREMENT
PLANE
63
Ideally, the XY-plane of the scanner has already been correctly aligned with
the sample plane using the three Levelling screws on the Scan head. How-
ever, this alignment can not be performed once the automatic approach has
been done, because this will damage the tip. During measurement, the
measurement plane should therefore be adjusted electronically using the
parameters ‘X-Slope’ and ‘Y-Slope’.
Maladjusted slope: measurement with improperly set X*-slope
Slope: Sample’s and measurement orientation before slope adjustment
X
Y
Z, Z*
Scanner XY-plane
'X-Slope' angle
'Y-Slope'
angle
'Rotation'
angle
Measurement plane
Image area
X*
Y*
Summary of Contents for easyScan 2 AFM
Page 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Page 7: ...7...
Page 86: ...AFM THEORY 86 Scanner coordinate system x y...
Page 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Page 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Page 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Page 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Page 164: ......