A
PPROACHING
THE
SAMPLE
51
The side view should look like figure
Side view of the cantilever after manual
. You can use the cantilever as a ruler to judge distances in
the views of the integrated optics.
Manual Coarse approach
In this step, the sample surface is brought within the range of the fine
approach stage.
- Use the three levelling screws (figure
(p.18)) to lower
the Scan head so that the cantilever is within 1-2 mm of the sample. Take
care that the Scan head remains levelled parallel to the sample surface by
turning all screws approximately the same amount.
When the sample is reflective, the mirror image of the cantilever should
now be visible (figure
Side view of the cantilever after manual coarse
). When the sample is not reflective, the shadow of the cantilever
may be visible. If neither a mirror image nor a shadow are visible, change
the light until it is visible, for example by adding a light source, or by shad-
ing the light from outside or from bright surfaces in the room.
Manual approach using the approach stage
In this step, the tip is brought as close to the sample surface as possible,
without running a risk of touching it. The closer the tip is, the less time the
automatic final approach takes.
- Observe the distance between tip and sample in the side view of the inte-
Side view of the cantilever after manual coarse approach
Summary of Contents for easyScan 2 AFM
Page 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Page 7: ...7...
Page 86: ...AFM THEORY 86 Scanner coordinate system x y...
Page 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Page 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Page 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Page 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Page 164: ......