P
REPARING
FOR
M
EASUREMENT
40
easyScan 2 controllerThe laser beam is now blocked by the DropStop. As a
consequence, the Probe Status light on the easyScan 2 controller will now
blink red.
- Place the cantilever insertion tool (figure
(p.17), 5) into the hole behind the alignment chip (figure
, top left
). The Cantilever Holder Spring opens.
- Use the Cantilever tweezers to remove the old cantilever from the instru-
ment (figure
).
It is recommended to always store the Scan head with a cantilever installed.
- Take the new cantilever out of its box. The cantilever is now oriented face
up.
Mounting the cantilever: top left: Inserting the cantilever insertion tool; top right: Inserting/
Removing the cantilever; bottom right: Correctly inserted cantilever