• images in the color map charts consist of uncorrelated lines only.
• images appear blurred.
- First follow the suggestions in section
Image quality suddenly deteriorates
(p.81). If these do not help, the cantilever should be changed.
If all peaks in the image have the same, usually triangular shape, the tip is
no longer sharp and has to be replaced. The measured structure is called a
tip artefact.
Adjusting the measurement plane
Ideally, the sample surface and the XY-plane of the scanner are all parallel.
But in most cases, the sample plane is tilted with respect to the XY-plane of
the scanner. In this case, the sample cross section in the X* measurement
direction has a certain slope. The Line graph chart in figure
is an example. This slope depends on the direction of the X* direction
and therefore on the rotation of the measurement, as shown in figure
.
This slope is undesirable for several reasons:
• It makes it difficult to see small details on the sample surface when the
Average, Plane fit or higher order filters can not be used.
• The Z-Controller functions less accurately, because it has to compensate
for the sample slope.
Measurement containing tip artefacts
Summary of Contents for easyScan 2 AFM
Page 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Page 7: ...7...
Page 86: ...AFM THEORY 86 Scanner coordinate system x y...
Page 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Page 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Page 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Page 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Page 164: ......