P
ROBLEMS
AND
S
OLUTIONS
80
• USB Video Adapter: Video Module version 1
• USB Composite device, USB 2.0 A/V Converter, USB EMP Audio
Device: Video Module version 2
• HID-compliant device, USB Human Interface Device: Nanosurf Analy-
sis/SPIP
• Aladdin HASP key, Aladdin USB key: Nanosurf Report
If there are problems with any of these drivers, or a wrong driver is installed,
you can try the following to fix it:
- Double click on the driver.
Properties dialog for the device now opens.
- Select the ‘Driver’-tab.
- Click the ‘Update Driver’-button
Windows will now ask you were to look for the driver.
- Tell windows to take the driver from the Installation CD.
AFM measurement problems
Probe Status light blinks red
This light will blink red when an insufficient amount of light reaches the
cantilever deflection detector. This can be either be due to the DropStop
still being closed or to a misaligned cantilever chip.
- Check that the cantilever is lying perfectly in the alignment chip.
- Remove the cantilever as described in section
(p.37). Blow away dust from the alignment chip and the cantilever’s back
side using dry, oil free air. Mount the cantilever again.
For further information, open the Scan Head Diagnosis dialog via the
menu ‘Options/Check Scan Head...’.
Summary of Contents for easyScan 2 AFM
Page 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Page 7: ...7...
Page 86: ...AFM THEORY 86 Scanner coordinate system x y...
Page 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Page 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Page 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Page 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Page 164: ......