Page 22 MSI-3750CS RF Weighmeter • User Guide
3750CS
C
ELL
S
CALE RF DIGITAL WEIGHT INDICATOR
®
SETUP UNITS
Enables the desired weight units for activation by the UNITS key. For example, if an operator only wants pounds
and kilograms activated, all other units can be disabled.
This operation requires that one of the Function Keys is setup as the UNITS key.
To Enable/Disable UNITS
UNITS
Allows easy weight units conversions. Up to 8 Units are available. See “SETUP UNITS” for details on
activating the available units.
This operation requires that one of the Function Keys is setup as the UNITS key. See “FUNCTION KEYS”. The example given below has F5 programmed
to be “UNITS” which is the default 3750CS mode. Change the example as required if one of the other keys is programmed to be “UNITS”. 3750CS
using Statistics are limited to one functional unit. Statistical Units can be changed only if all the ID Code Totals are Cleared using the “CLEAR ALL
TOTALS” procedure in Section 4.
To Change the UNITS
PEAK HOLD (FUNCTION KEY OPTION)
Allows monitoring peak weight. Uses a special mode of the CellScale’s A/D converter for capturing
transient events.
To Capture Peak Weight Readings
1) Program a
FUNCTION
key for “PEAKHOLD” (See Section 4).
2) Turn on the Peak Hold mode by pushing the selected
FUNCTION
key. The display reads “PEAK NET”
or “PK GROSS” to indicate peak hold mode is enabled. The display will only update when a greater
KILOGRAM
SHORTTON
1
2
UNITS
3
4
5
KILOGRAM
Scroll Through Available units
5
GRAMS
6
Push UNITS (Usually F5). The displayed weight
will change to the next activated unit. If more than
two units are activated, pushing the UNITS key will
rotate through the available choices.
1) Push the SETUP key
2) Push the UNITS key (usually F5). The menu will
display “UNITS”
Tons available on capacities 2,000 lb. and over. Metric Tons available
on capacities 1000 kg and over. Oz and g available on scales 250 lb
and under.
3) Push ENTER. The unit POUNDS will appear.
4) The numeric display will read “On” if the displayed
unit is activated or will read “OFF” if it is inactivated.
Use the LEFT or RIGHT Cursor keys to change the
status of the displayed unit from “On” to “Off”
5) Use the UP Scroll key to scroll through all the
available UNITS and use the LEFT or RIGHT
Cursor keys to enable (On) or disable (OFF) the
unit as desired.
If the user attempts to disable all units, Pounds will automatically be
enabled. OIML units are limited to kg, g, and metric Tons.
6) When finished selecting units, push EXIT to return
to normal scale operation.