48
6. Replicates. Set the number of scans.
7. Interval. Set the update interval of time scan. Update every 0.1 second.
8.PMT. Set the signal PMT level of time scan.
9.Most Data. Set the maximum sample size of time scan.
10.Multi-Curve. Set the display mode of curves. “Overlap” shows different curves in the
same window. “Overlay” will cover the last time scan curve, only showing the latest.
Fig.2-14 Time Scan Parameter Settings – Replicate Scan
11.Delay. Set the delay time between two scans.
12.Auto Saved Name. Edit the filename of spectrum. The software will automatically add the
current time of scan after the filename.
13.Auto Saved Direction. Set the autosave position in your computer.
8.1.3
Spectrum Processing
The functions and the operation are the same as those of spectrum processing in wavelength
scanning mode. For more details, refer to Section 7.2.2.
8.1.4
Storage Processing
The functions and the operation are the same as those of spectrum processing in wavelength
scanning mode. For more details, refer to Section 7.2.3.
8.1.5
Display Setting
The functions and the operation are the same as those of spectrum processing in wavelength
scanning mode. For more details, refer to Section 7.2.4.
INSTRUCTION MANUAL FOR SPECTRO 96 FLUOROSPECTROPHOTOMETER