47
8
Operation of Time Scanning Mode (When Connected with
a Computer)
8.1
Interface and Operation of Time Scanning Mode
Time scan differs from wavelength scan in that the unit of X-axis is time during scanning.
This working mode can be used to observe the situation in which the fluorescent intensity of a
sample varies with time at a fixed emission wavelength.
8.1.1
Interface of Time Scanning Mode
Click “Working Mode” in the menu bar in Fig.2-1. Then click “Time Scan” in the pull-down
sub-menu and the interface of wavelength can be changed to the interface of time scan. See Fig.2-3.
The time scan interface is the same as that of wavelength scan interface except for the unit of
X-axis. The functions of the shortcuts are also the same. For more details, please turn to Section 7.1.
8.1.2
Parameter Setting
Click “Parameter Settings” in Fig.2-3 to enter Time Scan Parameter Settings. As in Fig.2-13&
2-14.
Fig.2-13 Time Scan Parameter Settings - Instrument
1. Lambda. Set the emission wavelength of the time scan.
2. Scan Time. Set the total scan time.
3. Y Axis Min. Set the minimum of Y axis in the scan spectrum.
4. Y Axis Max. Set the maximum of Y axis in the scan spectrum.
5. Response. Set the signal integral time of time scan.
INSTRUCTION MANUAL FOR SPECTRO 96 FLUOROSPECTROPHOTOMETER