9.2.4
Standard Sample Setting
After clicking the “Set Standard.” shortcut on the bottom left of the interface of quantitative
analysis, a dialog box of “Set Standard Samples” shown in Fig.2-18 appears. Concentration unit
can be set in “Unit” box. If there are 5 standard samples and their concentrations are 10 for No.1
sample, 20 for No.2 sample, 30 for No.3…….respectively. After inputting “1” in the “Name” and
“10” in the “Concentration”, click the “Add” button. The data input in will be displayed in the box
below. Repeat the above operation for 5 times. Use the “Insert” and “Delete” buttons to edit your
sample data. The data will be displayed in the “Standard Samples” box in the interface of
quantitative analysis. See Fig.2-20.
Fig. 2-20 Dialog Box of Set Standard Samples
Note:
①
The parameter setting of standard samples must follow the order of concentration
from low to high.
②
The names of standard samples (including sample names) can be set for
Arabic numerals, English alphabet or Chinese characters.
9.2.5
Measurement of Standard Samples
In the “Standard Samples” in the interface of quantitative analysis, 5 rows of data of standard
samples input in are listed respectively (See Fig.2-19). When measuring, put the No.1 standard
sample into the sample compartment. Then double click the row of the No.1 standard sample. The
display of “Under Measurement” will appear on the bottom of the interface. A few seconds later,
fluorescence value (F) will be displayed. Repeat the above procedures until the fluorescence values
of all standard samples are measured out.