Installation of the Instrument
3.1
Installation Environment
The instrument is suitably used in a laboratory room to perform analytical measurements. Since
it works in combination with a computer, the working conditions must meet the following demands.
3.1.1
Space Environment of Laboratory
Environment temperature range:10
-
30
℃
. Environment humidity is less than 85
℅
.Corrosive
gas and other organic or inorganic solvent gas that is absorptive within the range of violet radiation
should not be contained in the atmosphere.
3.1.2
Working Table
The instrument should be placed on a sturdy and level table or the like, avoiding strong
vibration, direct sun light, and dust.
3.1.3
Power Source
Alternating voltage must be within a range of 220V
±
22V, 110V±11V.
3.1.4
Environmental Alternation
If the instrument needs to be used at a worksite, the working environment should accord with
the above demands for environment. If moving the instrument to other places, please use original
packing materials. If users have any special demands, may place a special order for the instrument.
INSTRUCTION MANUAL FOR SPECTRO 96 FLUOROSPECTROPHOTOMETER