Performance Index of the Instrument (Specifications)
Light source
150W Xenon lamp (consumable part)
Excitation filter
Filter character: ±5nm labeled
(An interference filter with center wavelength 365nm used as standard
configuration. Users may separately purchase other
φ
25mm
interference filters with center wavelength within 200
―
850nm range.)
Emission
monochromator
Emission monochromator wavelength range: (Em) 200
―
900nm
Wavelength accuracy within ±1nm
Wavelength repetition
≤
0.5nm
Emission spectrum bandwidth: 10nm
S
⁄
N ratio
S/N
≥
150 (Using 1cm quartz sample cell, measure the signal noise ratio
of Raman spectrum of water )
Check utmost
1×10
-10
g/ml sulfuric acid quinine solution
Measure linearity
≥
0.995
Peak value
≤
1.5%
Zero line drift
≤
0.3% (Within 10 minutes)
Upper limit change
of indicating value
±1.5% (Within 10 minutes)
Power requirement
220V±22V 110V/60HZ
Dimensions
442
×
392
×
250
(
mm
)
Net weight 11kg
Weight
Gross weight 13kg
Table 1-1 Performance Index of the Instrument