Operation of Quantitative Analysis
The interface of quantitative analysis is shown in Fig. 2-19. On the left is the display box of
Standard Sample and on the right is the display box of Unkown Sample.
9.2.1
Setting of Fluorescence Wavelength for Measurement
Click “Set Standard” below.Input the value of fluorescence wavelength of the sample to be
measured, then click “OK”.
Fig. 2-19 Interface of Quantitative Analysis
9.2.2
PMT Setting
There are 17 grades (1-17). Default PMT level is 1. Generally adjust PMT level by using the
highest concentration of the solution. Adjust the fluorescence value to make sure it is stable within
the range of 300-400.
9.2.3
Zero-point Adjustment
Place the blank solution into the light path. Click the “Adjust Zero” button (See Fig.2-19). At
this time the fluorescence value on the upper right of the interface becomes 0.0.
9. Operation of Quantitative Analysis Mode (When
connected with a Computer)
double click and measurement the
current standard samples