42
8.PMT Level. Set the PMT level of time scan.
9.Multi-Curve. Set the display mode of curves. “Overlap” shows different curves in the same
window. “Overlay” will cover the last time scan curve, only showing the latest.
10. Draw Mode
.
Set the plot mode of wavelength scan. “Real Time” will show the current
data of wavelength scan. “Batch” will show all data after the wavelength scan.
11. Slit
.
Its value is unchangeable and it has a bandwidth of 10nm.
Fig. 2-7 Interface of Wavelength Scan Parameter – Replicate Scan Settings
12.Delay. Set the Delay between two scans.
13.Autosave Filename. Edit the filename of spectrum. The software will automatically add
the current time of scan after the filename.
14.Autosave Position. Set the autosave position in your computer.
INSTRUCTION MANUAL FOR SPECTRO 96 FLUOROSPECTROPHOTOMETER