92
Resolution:
Use the BioMed SETUP menu Date and Time option to set the date and
time.
LCD Display:
***ALERT *** THERAPY TIME INVALID
Alert:
Invalid therapy time calculated.
Cause:
While programming or modifying an Intermittent therapy, the combination
of dose duration and dose frequency result in a prescription with a duration
greater than 288:00 HH:MM.
Resolution:
Recheck data entry. The changes to the field must comply with the total time
limit.
LCD Display:
***ALERT *** TIME LESS THAN RAMP TIME
Alert:
Unable to enter a TPN therapy time that is less than the ramp times.
Cause:
Time value entered for TPN infusion is less than ramp times.
Resolution:
Recheck data entry. TPN infusion time cannot be less than specified ramp
times.
LCD Display:
***ALERT *** VALUE PREVENTS A SUBSEQUENT DOSE
Alert:
Value entered will prevent the ability to program any more doses.
Cause:
The value entered for the “Variable Dose Amt” or “Time” field will cause no
remaining amount or time for a subsequent.
Resolution:
Recheck data entry.
If this is a new program, reprogramming of the therapy will be required if the
error is in a previously programmed dose.
If the therapy is in progress or this is a repeat program that has not started
and the error is in a previously programmed dose, turn the pump off and
then on. Choosing to resume or repeat the therapy will display the last
prescription that was completely programmed.
LCD Display:
***ALERT *** EMPTY LIBRARY
Alert:
Empty Library
Cause:
Occurs when “LIBRARY” is selected and no library is present.
Resolution:
Load a Library or program the pump using the “PROGRAM” option.
LCD Display:
***ALERT *** PUMP DISABLED CALL PROVIDER
Alert:
The pump is disabled
Cause:
The pump calibration data is invalid.
Resolution:
The pump needs to be returned for service.
LCD Display:
***ALERT*** DOOR OPEN or SET NOT PROPERLY INSTALLED
Alert:
Door is opened or administration set is not properly installed while the pump
is running
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...