If it is not necessary to preserve the therapy in progress or any other
programmed therapy, use the “CLEAR Rx” feature of the Clinician BioMed
menu. This feature will erase all programmed therapies in the pump. The date
and time can then be changed.
If the therapy is to be terminated but the prescription preserved (excluding
Delayed Start), go to the Pause menu, select “REPEAT Rx” and press YES at
the “Rx in progress” alert screen. The pump can then be turned off and then
on to access the Date/Time change option in the Clinician BioMed menu.
Selecting “NEW PROGRAM” and the name for the therapy in progress,
instead of “REPEAT Rx”, may also be used but this selection will erase the
prescription for the therapy.
LCD Display:
***ALERT *** DATE/TIME INVALID
Alert:
Unable to enter invalid date or time
Cause:
Date or time entry is not in a valid format.
Resolution:
Recheck data entry. An invalid date or time will not be accepted.
LCD Display:
***ALERT *** DELAY START TIME NOT VALID
Alert:
Delay Start Time not valid
Cause:
Delay start time entered is less than one minute or greater than 24 hours from
the current system time.
Resolution:
Reset Delay Start Time to fall within pump limits. (To obtain the current
date and time of the pump, turn the pump off and then on and observe the
time on the Welcome Screen.)
LCD Display:
***ALERT *** DOOR OPEN or SET NOT PROPERLY INSTALLED
Alert:
Door Open or Administration Set is not properly installed.
Cause:
Door Latch is not completely closed or the Administration Set is not installed
according to installation instructions.
Resolution:
Recheck the door latch assembly and securely close the door. If the door was
closed properly, re-install the set according to the installation instructions.
LCD Display:
***ALERT *** DOSE AMT IS LESS THAN DOSE INFUSED
Alert:
Dose amount entered is less than the amount already infused for the dose.
Cause:
New dose value entered is less than the amount already infused for the dose.
Resolution:
Recheck data entry. After the dose in progress finishes, the dose amount may
be changed to the lower value.
If the above is not suitable, re-program the prescription.
LCD Display:
***ALERT *** DOSE CANCELLED, RESUME INFUSION, THEN NEW
DOSE
T
roubleshooting
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...