90
Resolution:
Remove the remote bolus cord from the pump if connected. Refrain from
pressing the BOLUS key.
LCD Display:
***ALERT *** NUMBER OF DOSES INVALID
Alert:
Number of doses entered exceeds maximum for therapy
Cause:
The number of doses per bag in Intermittent therapy exceeds the allowable
maximum.
Resolution:
Recheck data entry. The number of doses requested must be less than or equal
to the maximum for the therapy.
LCD Display:
***ALERT *** PERFORM ROUTINE MAINTENANCE
Alert:
Maintenance Date (displayed only if date has arrived when pump is turned
on)
Cause:
Current Date is beyond the maintenance date.
Resolution:
Return pump for routine maintenance.
LCD Display:
***ALERT *** RAMP INVALID FOR AMT/RATE
Alert:
Ramp time entered is invalid for amt/rate already programmed
Cause:
Ramp value entered for TPN is unacceptable given the desired rate and
amount programmed.
Resolution:
Recheck all data entered for prescription. If programming a new TPN therapy
(“NEW PROGRAM” selected) and the ramp time is a requirement, enter the
ramp time(s) and total time first followed by the next governing parameter
(rate or volume to be infused).
LCD Display:
***ALERT *** RATE EXCEEDS DELTA RATE/DELTA TIME
Alert:
Rate entered exceeds the delta time limits
Cause:
The rate entered falls outside of the allowable range for the delta rate / delta
time limits programmed.
Resolution:
Recheck data entry or wait more time elapses then attempt to change the rate.
Rate selected cannot exceed the range placed on medLMTS
™
.
LCD Display:
***ALERT *** RUNNING, PAUSE PUMP PRIOR TO PRESSING OFF
Alert:
Pause Infusion before turning off
Cause:
The OFF key was pressed while the pump was infusing or a running screen
displayed.
Resolution:
Press PAUSE key before pressing OFF key to stop an infusion.
LCD Display:
***ALERT *** Rx IN PROGRESS, YES to Repeat – NO to Return
Alert:
“REPEAT Rx” selected while a therapy is already in progress
Cause:
“REPEAT Rx” was selected and there is a therapy currently in progress.
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...