Alert:
Cancellation of a clinician dose is pending and can only be processed with
the pump running.
Cause:
Clinician dose in progress is canceled while the infusion is paused; and the
user attempts to enter a new Clinician Dose, prior to resuming the infusion
Resolution:
Resume the infusion and then program the desired Clinician Dose.
LCD Display:
***ALERT *** DOSE FREQ LESS THAN DOSE DURATION
Alert:
Dose frequency time is less than the time required to infuse the programmed
dose.
Cause:
The dose frequency value entered is less than the time required to infuse the
programmed dose amount.
Resolution:
Recheck data entry. Dose frequency cannot be less than required time for
dose to infuse.
LCD Display:
***ALERT *** EMPTY BAG
Alert:
Empty Bag
Cause:
Occurs when the volume infused for the prescription is equal to the “Bag
Vol.” Bag volume may have been reduced due to priming.
Resolution:
Verify that the volume remaining in the source container is sufficient to
complete the current therapy. If the volume is insufficient, add another source
container or stop the infusion.
LCD Display:
***ALERT *** EXCEEDS POSSIBLE DOSES PER HR
Alert:
Unable to enter # of boluses/hr that would be possible given the programmed
bolus amount.
Cause:
# Boluses/hr value exceeds the possible # given the bolus amount.
Resolution:
Recheck data entered. The # of boluses per hour cannot exceed the time
required to deliver the bolus amount programmed.
LCD Display:
***ALERT *** EXTERNAL POWER DISCONNECTED
Alert:
External power source disconnected.
Cause:
External power source is disconnected or external battery pack power is
completely depleted during pump operation.
Resolution:
Reestablish external power source, or let pump continue to operate on
batteries.
Note:
The battery level should be checked after the external source is
disconnected to gage if additional batteries will be required before an
external power source is available again.
LCD Display:
***ALERT *** FIELD NOT ACCEPTED
Alert:
Field Not Accepted.
Cause:
The field indicated by the flashing cursor has not been accepted.
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...