This section assists in identifying and resolving alert and alarm conditions that may occur during the
use of the
Moog Curlin Infusion
pump.
Alerts, Alarms, and Error Codes
Alerts, Alarms, and Error Codes are classes of events that are relayed
to the user and logged into the history file. The screen will display
messages, the LED status lights will blink, and the audio alarm will
sound. The backlight feature will illuminate the screen during an
alert or alarm.
If the batteries have been dislodged or are so low in power that they
cannot operate the pump, a solid tone may occur accompanied by
a blank display screen and a constant red alarm light. In this event,
replace the batteries.
ALERTS
Alerts report conditions that require attention from the user but may not be safety issues. When
an alert occurs, a message is displayed on the LCD screen, and the audio alarm is sounded. The
message displays on the screen for three seconds if it is a user-initiated alert, such as a wrong key
press. If it is an automatic alert occurring while a therapy is infusing, such as “low battery,” the
message displays for twenty seconds. If it is an alert that requires a user response, the alert remains
displayed until the user presses one of the keys specified on the screen. The following is a list of the
alerts in alphabetical order.
LCD Display:
***ALERT *** AIR Sensor OFF. Using In Line Air Filter?
Alert:
Air-In-Line Sensor Disabled
Cause:
The Air-In-Line Sensor has been disabled and either the RUN key has been
pressed to run an infusion or the AIR SENS setting has been set to OFF on
the Options menu while the pump is running.
Resolution:
If an air-eliminating filter is being used, press YES. The infusion will start/
resume and the No Air-In-Line sensitivity will be disabled until the empty bag
alert occurs or infusion complete is reached. No Air-In-Line alarms will
occur while the air sensitivity is disabled.
If no air-eliminating filter is being used, press NO. The infusion will start/
resume and the Air-In-Line sensitivity will be set to 2 mL for the
IMPORTANT
INFORMATION
T
roubleshooting
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...